Flight DEFENDME Aerosol Evacuator R.1.2
12
Secondary Filter Assembly
3-6 months
Displays 050 or a multiple of
it.
Eg.050/100/150
Main Filter Assembly
12 Months
Display 100 or a multiple of it
Eg. 100/200/300
Replace relevant filter according to the duration (months) or if the flow value has
been reached. Or when the LED display shows a circle than the filter must be replaced.
The flow value is a three-digit value; the cumulative value is 001 when the fan rotates 5.76
million revolutions.
7.2.2 How to Replace Filters
Release the four latches on the upper cover of the machine and remove the top cover. Inside
you will see the primary microfiber filter, secondary filter assembly and main filter in that order.
Replace each filter accordingly. When replacing the main filter ensure that the filter is placed
correctly and that the plug for the fan is connected to the bottom of the filter.
8.0
Troubleshooting
If the device malfunctions please follow the table below before contacting your dealer to resolve
the issue. If the problem persists please contact your local authorized distributor or dealer.
Problem
Reason
Solution
The display flashes “OFF” and
a circle rotates clockwise on
the display or the alarm goes
off and the fan stops working
The air inlet is completely
blocked.
Turn off the power and check
the directional air duct and
suction nozzle to see if there
is a blockage. Remove any
blockages.
If the problem persists, open
up the upper cover and check
When replacing the filter ensure that the power is turned off and the unit is unplugged.