Entering Output Options 

The next step in programming your F601 will be programming the data display and storage details 

under 

Output Options

10)

 

Using the 

or 

keys, place the carrots around 

>OPT<

 

(

Output Options

) then press 

Enter

 

11)

 

The following screen offers you the choice of 

Channel A

 

or 

Channel B

. Using the

 

and 

keys will change the 

Channel letter. When the correct channel is displayed, 
(which is typically Channel A) press 

ENTER

 

12)

 

Using the

 

and 

keys to change the Type of 

flow you’d 

like to measure. Your Options are 

Volume Flow, 

Velocity Flow, Mass Flow and Heat Flow

. Press 

ENTER

 

 

13)

 

Using the

 

and 

keys to change the units of flow here. 

Your Options are numerous, in 

Metric 

and 

Imperial

 

units. 

 

14)

 

All displayed data will be applied to a 

Floating Average 

Function to remove minor spikes in data. Range 1 to 100 
seconds. 

 

15)

 

Use the 

or 

keys to make your data storeage selection. 

Choosing 

>YES< 

will initiate the data logging function 

that will be defined later. 

 

16)

 

Use the 

or 

keys to activate 

RS232 Serial Output

Unless you have specific need for this feature, we 
recommend to leave this off by choosing 

>NO<

 

 
 

17)

 

Use the

 

and 

keys to select your data storeage rate. 

 
 

 

That completes the Output Options programming. 

You will be returned to the Main Menu 

Follow the steps on the next page to begin gathering data. 

  

  par        mea         >OPT<          sf 

 

Output Options 

 

Output Options 
For Channel                  A: 

 

Physical Quantities 
Volume Flow                 

 

Volume In: 
USgpm                 

 

Damping 
                10                   sec

 

 

Store Meas. Data 
no                             >YES< 

 

Serial Output 
>NO<                          yes 

 

Storage Rate 
     Once per 10 sec.

 