
18 Settings
FLUXUS F501
18.3 Libraries
125
UMFLUXUS_F501V1-3EN, 2020-06-30
It is possible to set a time period after which the display toggles between the positive and the negative totalizer during the
measurement.
• Enter a time interval between 0 (off) and 5 s.
• Press ENTER.
18.3
Libraries
The internal material database of the transmitter contains parameters for pipe and lining materials as well as for fluids.
The material and fluid scroll list displayed in the program branch
Parameter
can be arranged. Shorter scroll lists make
work more effective.
• Select the menu item
Special Funct.\SYSTEM settings\Libraries
.
• Press ENTER.
• Select
Material list
to edit the material scroll list or
Medium list
to edit the fluid scroll list.
• Select
go back
to return to the menu item
SYSTEM settings
.
• Press ENTER.
• Select
factory
if all materials/fluids of the internal database are to be displayed in the scroll list. An already existing
scroll list will not be deleted but only deactivated.
• Select
user
to activate the user defined scroll list.
• Press ENTER.
If
user
is selected, the material or fluid scroll list can be edited, see section 18.3.1...18.3.3.
• Select
End of Edit
to stop editing.
• Press ENTER.
• Select
yes
to store all changes of the scroll list or
no
to quit the menu item without storing.
• Press ENTER.
Special Funct.\...\Measuring\Toggle totalizer
Special Funct.\SYSTEM settings\Libraries
Special Funct.\SYSTEM settings\Libraries\Material list
Special Funct.\...\Material list\user\Show list
Special Funct.\...\Material list\user\End of Edit
Special Funct.\...\Material list\user\Save List?
Notice!
If the material/fluid scroll list is quit by pressing key BRK before storing, all changes will be lost.