17 SuperUser mode
17.1 Activation/deactivation
FLUXUS F501
2020-06-30, UMFLUXUS_F501V1-3EN
114
17
SuperUser mode
The SuperUser mode offers the possibility of an advanced diagnostic of signals and measured values as well as the definition
of additional parameters adapted to the measuring point, in order to achieve better measured values or for experimental work.
Special features of the SuperUser mode are:
• Default settings will not be used.
• There are no plausibility tests during the parameter entry.
• It is not checked whether the entered parameters are within the limits given by physical laws and technical data.
• The cut-off flow is not activated.
• A number of sound paths has to be entered.
Some menu items that are not visible in the normal mode are displayed.
17.1
Activation/deactivation
• Enter the HotCode
071049
, see section 8.4.
• Press ENTER.
It is displayed that the SuperUser mode is activated.
• Press ENTER. The main menu will be displayed.
• Enter the HotCode
071049
again to deactivate the SuperUser mode.
It is displayed that the SuperUser mode is deactivated.
• Press ENTER. The main menu will be displayed.
17.2
Defining flow parameters
In the SuperUser mode, it is possible to define some flow parameters (profile bounds, correction of the flow velocity) for
the specific application or measuring point.
• Select the menu item
Special Funct.
\
SYSTEM settings\Measuring\Calibration
.
• Press ENTER.
17.2.1
Profile bounds
• Select
user
to define the profile bounds. If
factory
is selected, the default profile bounds will be used and the menu
item
Calibration
will be displayed.
• Press ENTER.
Notice!
The SuperUser mode is intended for experienced users with advanced application knowledge. The parameters can affect
the normal measuring mode and lead to wrong measured values or to a measurement failure when a new measuring point
is set up.
SUPERUSER MODE\IS ACTIVE NOW
SUPERUSER MODE\IS PASSIVE NOW
Notice!
Some of the defined parameters are still activated after the deactivation of the SuperUser mode.
Special Funct.\SYSTEM settings\Measuring\Calibration
Special Funct.\...\Calibration\...\Profile bounds