28
6 - CLEANING, SANITISATION, DISINFECTION
Before and after each use, the device and the accessories must be cleaned
and disinfected properly as described below. If this is not done, some mi-
croorganisms may collect in the device, causing the risk of infection. Abso-
lutely do not use alcohol to clean the device
6.1. CLEANING Accessories and Removable Upper Part (2):
6.1.1. WASHING:
Wash the components
2-2G-3-4-5
with lukewarm drinking water with mild deter-
gent for dishes (non-abrasive), then rinse thor-
oughly with a jet of lukewarm drinking water to
remove any detergent residue.
6.1.2. DRYING
6.1.2.a.
After washing and rinsing all the compo-
nents, shake them to remove excess water and
reassemble them, as described in the following
paragraph
7. REASSEMBLY OF THE DEVICE
. At
this point, to eliminate the residual water that
has deposited in the
nebuliser head (2G)
, press
Button 1A
for about 2-3 seconds until the de-
vice stops spraying
(side figure).
This operation is
very important, because it prevents the deposit
of limestone in the micropores of the
Nebuliser
Head (2G)
which could affect proper nebulisation
of the drug in subsequent therapeutic applica-
tions.
6.1.2.b.
To complete the drying of all the compo-
nents, leave them out and in a dry place (e.g. not
in the bathroom) as into the side figure.
WARNING
Do not clean the
Mesh (2G)
using brushes or
other instruments: this may cause irreparable
damage.
2G
2
3
5
4
2G
2
3
5
4
2G
2
3
5
4
5
2
2G
3
4
PUSH