FORCE PLATE

 

 

400 Series 

Performance Force 

Plate 

Summary of Contents for 400 Series

Page 1: ...Technology 21 Bishop St Skye SA 5072 Australia 61 8 8331 9229 Mob 0418 815 400 www fittech com au Copyright Fitness Technology Adelaide Australia 21 st March 2005 FORCE PLATE 400 Series Performance Force Plate ...

Page 2: ...able of running the BMS LPT auto saving all data files outputting to Excel via the BMS software program 6 Only Performance Force Plate on the market capable of running on our new Inner Balance software tracking used to measure record all Anterior Posterior Medial Linear movements 7 400S Weight 20 6 Kg 45 3 lbs 8 1 x PCB Module operates 4 Load Cells LPT powered via 5 V DC Computer USB Port only 9 1...

Page 3: ... and White wires maybe reversed Ex Black Shield Heat shrink wire colour may vary Installation 1 Carefully unpack the platform by lifting it out the end of the transport carton Make sure all items are present Report any discrepancies immediately 2 Place the platform onto a flat stable surface 3 Install each foot into the load cell holes on the four corners of the platform by rotating the feet to th...

Page 4: ...18 815 400 www fittech com au Copyright Fitness Technology Adelaide Australia 21 st March 2005 Issue 2 00 21st Aug 2008 USB Cable Connections USB Connector Lead to RJ45 Connector to External RJ45 Socket for LPT cable connection USB Connector US8 8 Plug w lead to RJ45 Skt Pin 2 PT5A IK Pot V Signal Pin 4 USB 5V Pin 6 USB Gnd Pin 7 MBU V Signal ...

Page 5: ...form cannot be used until the 4 x feet with M12 1 75 pitch threads are installed in each of the four corners They screw into the Platform with standard a right hand thread Stainless Steel M12 spacer Nuts are used for levelling and adjusting the height of the Platform If required the M12 spacer nuts can be removed but if this is done never thread the 4 x feet threads thru all the way back off 5mm R...

Page 6: ...a 21 st March 2005 Issue 2 00 21st Aug 2008 1 of the 4 x low profile shear beams bolted in position Care and Maintenance of Force Plate To clean the Force Plate we recommend the use of standard household detergent to wipe down the external surface area Note Do not place or allow any water to ingress into the Platform via the load cell holes or the penetration used for the USB communication port un...

Page 7: ...per use appear therein and arise solely from faulty materials or workmanship will be remedied by us free of charge provided the equipment is returned to our Workshop within a period of twelve 12 months from date of delivery freight paid both ways In the case of warranty service to equipment on site or at the purchaser s premises all traveling and accommodation costs shall be to the purchaser s acc...

Page 8: ...t you disconnect any USB extension leads 4 Now if the 400S is still not communicating with your PC we have made available a test program which allows the XPV7 circuit board inside the 400S to be tested When performing these tests it is best to remove the cover plate from the underside of the platform to expose the XPV7 circuit board The test program is available from http www innervations com down...

Page 9: ... correctly you should see the version and serial numbers for the XPV7 in the title line of the test program Further confirmation that the XPV7 is communicating with your PC is the PCB System Information and PCB A D Settings which are shown at the bottom of the test program dialogue box If this information is not displayed then there is either still a problem with your PC or the XPV7 is not running...

Page 10: ...you will see analogue bars next to each channel and a digital value on the right hand side If nothing is displayed this is because the internal error checking in the XPV7 has detected a fault and is not sending faulty data 8 If you are not receiving data then press the Stop Continuous Data Transfer button 9 So that you can determine which channel has caused the fault you must click on the DISABLE ...

Page 11: ...gged into channel 3 4 and 3 4 go into 1 2 Now reconnect the USB cable and run the test program and repeat step 8 If the faulty channel moves in this case to CH 1 then there is a connection cable or load cell fault If on the other hand the fault doesn t move then the XPV7 circuit board is faulty 11 Next you should compare the A D Settings at the bottom of the dialogue box these numbers should match...