March 2005. Issue 2.00, 21st Aug 2008
10.
To determine the cause of the fault close the test program, disconnect the USB cable, and swap
over the pluggable terminal blocks so that the channel 1 & 2 pair is plugged into channel 3 & 4 and 3
& 4 go into 1 & 2. Now reconnect the USB cable and run the test program and repeat step 8. If the
faulty channel moves (in this case to CH #1) then there is a connection, cable, or load cell fault. If on
the other hand the fault doesn’t move then the XPV7 circuit board is faulty.
11.
Next you should compare the A/D Settings at the bottom of the dialogue box, these numbers should
match exactly with what you have on screen. If these numbers do match then there is a hardware
fault on the XPV7 circuit board.
If these numbers don’t match then contact our repair centre for further instructions on resetting these
values.