-Flowmeter
Cleaning
With the conclusion of the flow measurements, the Flowmeter is to be thoroughly cleaned.
Follow these steps:
-
Ensure that the Flowmeter is not under pressure by making certain that hand valve (Pos.
8) is in OFF-position. Make sure reading on pressure gage is nill, then remove
compressed air connection.
-
Unscrew all manually screwed parts (nozzle, nozzle carrier and measuring cylinder),
remove pressure plate.
-
Use only cleaning agents that are compatible with the product measured and which will
not harm any of the metal parts, especially the pressure plate.
-
Clean the threads of the parts, including the pressure cylinder, thoroughly. It is important
however, that the threads are greased again prior to remounting the parts. Use only
grease which does not contain any parts of Silicon and which will not react with products
to be measured. Remount cleaned and greased parts as indicated under 'Preparation for
measurements'.
Your Flowmeter is now ready for further measurements, store in a well protected and clean
spot.