Page 32 of 34
SUPER SESI, by FIT
User manual, Revision 2.0
Resetting overheating protection relays
The sample line and the ionization chamber are protected against overheating with resettable
thermal relays. If the overheating relay is activated, it will shut the corresponding heater down
until it is reset by the user. In order to reset them, turn of the main switch, wait until the system
cools down, and restart the power. If the problem persists, contact FIT for technical assistance.
Replacing the fuses
In order to change a fuse, first make sure that the power cord is not connected. Open the fuse
holder, and replace the fuses. The fuse type is
‘
Quick Acting F LBC Fuse, 1A, 5x20 mm
’
.
Trouble-shooting
SUPER SESI is provided with two years of technical service. Please contact us via email or
telephone if you need further assistance. In order to allow us to give you a better service, please
specify the type of problem and we will get back to you as soon as possible.