EasyGrid LT User Guide
20
Analog Outputs
4.11
Each analog output offers a standard 4-20 mA factory configuration.
Analog outputs provide a means of transmitting temperature
data directly from the unit to an external data-gathering
system or monitoring equipment. The EasyGrid LT provides
one analog output for each temperature channel present in
the unit.
Wire length and gage are related when connecting analog
outputs. For a run of 100 m or less, recommended gage is 24
AWG or greater. The longer the run, the smaller the
recommended gage will be. Recommendations call for
shielded wire, or for installation in a metal conduit.
Please refer to Channel Setup and Analog Output Scale section of the Nortech Client User Guide for
detailed information.
Communication Configuration
4.12
4.12.1
USB
The USB configuration port is accessible under the protection door on the left
front panel.
The USB port is used to link to a PC for parameter configuration, data retrieval and
firmware update using the Client Software.