Auto-tuning (PID
Parameter Calibration)
Step 4: Terminate configuration mode as described in Configuration
Mode. To cancel the auto-tuning during the process, turn off
the main power switch. If cancelled, the Optimal Mode will be
selected automatically with an alarm.
Coefficient (COEF)
Five user-selectable temperature control modes are provided for your
convenience.
Step 1: Press the CONFIG Button.
Step 2: Select temperature control mode selection [CoEF] shown at
right by turning the Stirrer Knob.
Step 3: Turn the Stirrer Knob to check the five temperature control
modes in order. And select the desired mode by using the
Stirrer Knob as shown below.