Check Configuration
Settings (LISt)
Step 5: Terminate configuration mode as described in Configuration
Mode.
Auto-tuning (PID
Parameter Calibration)
The PID parameters for temperature control can be automatically tuned
to take your specific operating circumstances into consideration.
If you select User Mode [Coefficient], the unit will be operated with the
auto tuning value.
The auto-tuning (calibration) procedures are as follows:
Step 1: Press the CONFIG Button.
Step 2: Select the auto-tuning mode [CALi] shown below by turning the
Stirrer Knob.
Step 3: Press the Stirrer Knob and select YES by turning the Stirrer
Knob. Press the Stirrer Knob again to start the auto-tuning
4-24
Section 4
Control Panel