The instrument will stop running in 2 minutes and alarm if
the normal operating is obstructed or the instrument is
malfunctioning.
Alarm Instructions
3.4 Tilt Angle Adjustment
Power Recovery
If the power supply is cut off suddenly while the
instrument is in operation, the unit will automatically run
at the previously set parameter upon power restoration.
The display window will flash. Press any button to stop
flashing.
Finish Operation
Note:
To ensure shaking operation smooth and steady, it may take
1minute for the microprocessor control system to
accelerate the tray to the set speed.
After the operation is finished, press the power switch at the
back right side the instrument and put it into the “O” state.
Unplug the instrument and store the instrument according
to the storage guide.
Operation Steps
1. Make sure the instrument is not running while adjusting
the tilt angle. Press the Stop button to stop the instrument
if it is in running.
2. Push the knob on the cover plate from “Run” to “Adjust
Angle”position as shown in figure 1.
3. Take off the dimpled rubber mat and put it aside.
4. Hold the tray with one hand and unscrew the inner
hexagon screw at the center of the tray with the angle
adjustment wrench to adjust the angle of the tray with the
other hand, as shown in figure 2.
5. Apply upward force at the front of the tray and
downward force at the back of the tray to adjust the tray
to the desired angle, as shown in figure 3.
6. Hold the tray with one hand and tighten the inner
hexagon screw with the allen wrench at the center of the
tray until it's fastened, as shown in figure 4.
7. Push the knob on the cover plate from “Adjust Angle” to
“Run” position as shown in figure 5.
8. Put on the dimpled rubber mat and complete the angle
adjustment, as shown in figure 6.
9. Follow steps on Section 3 to operate the equipment.
Warning:
1.The instrument should be unloaded and stopped at the
specified position when adjusting the angle. It is not
allowed to adjust the angle at other positions.
2.The instrument can be started only when the knob is at
the “Run” position after the angle adjustment. Otherwise
the instrument may be damaged.
Figure
1
Figure 3
Figure 5
Figure 2
Figure 4
Figure 6
F
F
Now you can start using this Wave Motion Shaker.