Fisher Scientific Fisherbrand 88861027 Operation Manual Download Page 8

7

The front panel of the Wave Motion Shaker contains all the controls needed to operate the unit.

TIME

This chapter covers the control panel and its operation.

Section 3 Operation 

3.1 Control Panel

1. Start/Stop button: Start or stop the instrument.
2. RUN indicator light: The light is on when the instrument is
running and off when the instrument is in standby.
3. TIME display window: The window shows cumulative time
(in continuous mode) or remaining time (in timer mode). The
range of time displayed is 0 to 99 hours and 59 minutes. The
accuracy is 1 minute. 
4. SPEED display window: The window shows set speed
(when the instrument is in standby) or current speed (when the
instrument is running). 
5. Set Time Buttons: UP/DOWN Arrow buttons are used to
increase/decrease the set time of the instrument.

. Set Speed Buttons: UP/DOWN Arrow buttons are used to

increase/ decrease the set speed of the instrument.

6

1. Connect all the components according to the figures shown
on page 4 of this manual. Use grounded power outlet.
2. Press down the power switch on the back right side of the
instrument and put it to the “|” state and then the instrument is
in standby.

3.2 Installation

3.3 Settings

Time Settings

1. Continuous mode

Press the “     ” or “     ” arrow button below the TIME display
window. When the number shown on the display window starts
flashing, press “     ” arrow button to decrease the time to 00:00
and then release the button. The time setting is finished after the
number shown on the display window has flashed twice.

2. Timer mode

window. When the number shown on the display window starts
flashing, press “     ” or “     ” arrow button to increase or
decrease the time value. Release the button when the time
shown on the display window reaches the set value. The time
setting is finished after the number shown on the display window
has flashed twice.

Speed Settings

Press the “     ” or “     ” arrow button below the SPEED display
window. When the number shown on the display window starts
flashing, press “     ” or “     ” arrow button to increase or
decrease the speed value. Release the button when the speed
shown on the display window reaches the set value. The speed
setting is finished after the number shown on the display window
has flashed twice 

Note: 

press the “     ” or “     ” arrow button for a longer time to

accelerate the setting.

Run and Stop

1. Continuous Mode

Press

button and the instrument will start running with the

specified settings and the RUN indicator light will be on. The
TIME display window will show the cumulative time and the
SPEED display window will show the current speed. Press “     ”
button again and the instrument will slow down until it stops. 
The instrument will then be in standby and the two display
windows will show the set values.

“     ”

2. Timer Mode

Press 

 button and the instrumentwill

The

TIME display window will show the remaining time and the
SPEED display window will show the current speed. Press “     ”
button again and the instrument will slow down until it stops.
The instrument will then be in standby and the two display
windows will show the set values.

“     ”

 start running with the

specified settings and the RUN indicator light will be on. 

Press the “     ” or “     ” arrow button below the TIME display

Summary of Contents for Fisherbrand 88861027

Page 1: ...Wave Motion Shaker Operations Manual V1 12 21 17...

Page 2: ...This manual covers the model shown below 1 NA Model Voltage Description 88861027 88861028 100 240V 100 240V Wave Motion Shaker US plug Wave Motion Shaker Intl plug EU Model N A 15524070...

Page 3: ...transmitted in any form or by any means including photocopying recording or using information storage and retrieval systems for any purpose other than the purchaser s own use without the express writt...

Page 4: ...onditions Safety Instructions Relationships between Speed and Load Section 3 Operation Control Panel Settings Tilt Angle Adjustment Section 4 Safety Tips and Maintenance Section 5 Troubleshooting Sect...

Page 5: ...Packing List 1 10 N A N A N A Wave Motion Shaker 6 2 1 1 10 1 N A 6 2 1 1 1 1 1 1 Figure 1 2 Connections Main Body Power Cable all the components before discarding the packing If there are any missing...

Page 6: ...read the entire instruction manual before operating the Wave Motion Shaker 2 3 Safety Instructions 5 Note Maximum speed may vary with heavy or unbalanced loads Rotation Speed Maximum Load Centered on...

Page 7: ...cement is needed please make sure to gradually increase the speed Tilt Angle 12 Load Weight kg Placement of Load Tilt Angle Centered Symmetrical 9 8 9 8 120 120 0 0 Rotation Speed rpm 80 80 Load Weigh...

Page 8: ...ndow has flashed twice 2 Timer mode window When the number shown on the display window starts flashing press or arrow button to increase or decrease the time value Release the button when the time sho...

Page 9: ...wn in figure 1 3 Take off the dimpled rubber mat and put it aside 4 Hold the tray with one hand and unscrew the inner hexagon screw at the center of the tray with the angle adjustment wrench to adjust...

Page 10: ...of the tray and put back the dimpled rubber mat onto the tray After prolonged use the small rubber tubes may be damaged Please follow the steps below to replace these small rubber tubes 1 Remove the...

Page 11: ...the instrument please shut down the instrument and clean up the liquid immediately If the liquid has already spilled into the unit cut off the power supply first and immediately clean up the liquid a...

Page 12: ...weighted or unbalanced load Electrical malfunction Mechanical malfunction Tray loose Keep record for maintenance Solution Connect the power Turn power switch on Replace power adaptor Adjust the weight...

Page 13: ...Spare Part 88861178 88861199 180 7mm 6pcs pack M4 8 pcs pack 8 Power Cable Spare Part US Plug Power Cable CN Spare Part Plug Power Cable EU Spare Part Plug Power Cable UK Spare Part Plug Screw for Rub...

Page 14: ...and maintenance this product is warranted for 24 months against defective materials or workmanship The 24 month warranty period begins from the delivery date of this product For product quality or per...

Page 15: ...478 it fishersci com Finland 358 0 9 8027 6280 fishersci fi France 33 0 388 67 14 14 fishersci fr North America Canada 1 800 234 7437 fishersci ca Netherlands 31 0 20 4887 70 00 nl fishersci com Norw...

Reviews: