2.1 Specifications

Section 2 Overview

WARNING DO NOT

 use the Wave Motion

Shaker in a hazardous atmosphere or with 
hazardous materials for which the unit was
not designed.  Also, the user should be
aware that the protection provided by the
equipment may be impaired if accessories
used are no provided or recommended by
the manufacturer, or are used in a manner
not specified by the manufacturer.

CAUTION ! 

To avoid electrical shock,

completely cut off power to the unit by
disconnecting the power cord from the unit
or unplug from the wall outlet. Disconnect
unit from the power supply prior to
maintenance and service. Any spills should
be removed promptly. Bio hazard spills
should be cleaned using approved liquid
promptly. Solvent spills are a fire hazard.
Stop the unit immediately, and DO NOT
operate until clean up is complete and
vapors have dissipated.

DO NOT 

immerse the unit for cleaning.

DO NOT 

operate the unit if it shows signs of

electrical or mechanical damage.

Please read the entire instruction manual before
operating the Wave Motion Shaker.

2.3 Safety Instructions

5

Note:

 Maximum speed may vary with heavy or unbalanced loads

.

Rotation Speed

Maximum Load

(

Centered on tray

)

9.8 kg (22Ibs)

5~120rpm

±

1rpm

3D Circular Motion

0~12

°

Adjustable

Speed Range

Speed Accuracy

Operation Mode

Carousel Angle Range

Display Accuracy

1rpm

1

min 99h

5

9min

~

LED

1

 

min

Timing Range

Display

Display Accuracy

Time

415 293 216mm(16.3 11.5 8.5in.)

×

×

×

×

316 276

mm(12.4 10.9

in.)

×

×19

×

×0.7

536 501 346mm(21.1 19.7 13.6in.)

×

×

×

×

Overall Dimensions

Tray Dimensions

Packaging Dimensions

Size

Net Weight

14Kg(30.8lbs)

16Kg(35.2lbs)

Gross Weight

Weight

AC100-240V, 50Hz/60Hz, 0.2A

Requirement

Power Supply

Certification

ROHS, WEEE, cCSAus, CE Mark

55dB with no load

Noise Level

Others

2.2 Environmental Conditions

Application Environmental Conditions: indoor use

Storage Environmental Conditions

5  to 40

℃(

41 to 104

℉)

Temperature

Humidity

Altitude

Voltage Fluctuation

±

10% of the nominal voltage

2,000 m

Temperature

Humidity

20

to 90

 non-condensing

%,

31

℃,

Relative

 

Humidity

80%, non-condensing

NA Cat

.

No.

NA/EU Cat

.

No.

Description 

Category Number

Wave Motion Shaker

88861027

US plug

88861028

/15524070

Intl plug

Summary of Contents for Fisherbrand 88861027

Page 1: ...Wave Motion Shaker Operations Manual V1 12 21 17...

Page 2: ...This manual covers the model shown below 1 NA Model Voltage Description 88861027 88861028 100 240V 100 240V Wave Motion Shaker US plug Wave Motion Shaker Intl plug EU Model N A 15524070...

Page 3: ...transmitted in any form or by any means including photocopying recording or using information storage and retrieval systems for any purpose other than the purchaser s own use without the express writt...

Page 4: ...onditions Safety Instructions Relationships between Speed and Load Section 3 Operation Control Panel Settings Tilt Angle Adjustment Section 4 Safety Tips and Maintenance Section 5 Troubleshooting Sect...

Page 5: ...Packing List 1 10 N A N A N A Wave Motion Shaker 6 2 1 1 10 1 N A 6 2 1 1 1 1 1 1 Figure 1 2 Connections Main Body Power Cable all the components before discarding the packing If there are any missing...

Page 6: ...read the entire instruction manual before operating the Wave Motion Shaker 2 3 Safety Instructions 5 Note Maximum speed may vary with heavy or unbalanced loads Rotation Speed Maximum Load Centered on...

Page 7: ...cement is needed please make sure to gradually increase the speed Tilt Angle 12 Load Weight kg Placement of Load Tilt Angle Centered Symmetrical 9 8 9 8 120 120 0 0 Rotation Speed rpm 80 80 Load Weigh...

Page 8: ...ndow has flashed twice 2 Timer mode window When the number shown on the display window starts flashing press or arrow button to increase or decrease the time value Release the button when the time sho...

Page 9: ...wn in figure 1 3 Take off the dimpled rubber mat and put it aside 4 Hold the tray with one hand and unscrew the inner hexagon screw at the center of the tray with the angle adjustment wrench to adjust...

Page 10: ...of the tray and put back the dimpled rubber mat onto the tray After prolonged use the small rubber tubes may be damaged Please follow the steps below to replace these small rubber tubes 1 Remove the...

Page 11: ...the instrument please shut down the instrument and clean up the liquid immediately If the liquid has already spilled into the unit cut off the power supply first and immediately clean up the liquid a...

Page 12: ...weighted or unbalanced load Electrical malfunction Mechanical malfunction Tray loose Keep record for maintenance Solution Connect the power Turn power switch on Replace power adaptor Adjust the weight...

Page 13: ...Spare Part 88861178 88861199 180 7mm 6pcs pack M4 8 pcs pack 8 Power Cable Spare Part US Plug Power Cable CN Spare Part Plug Power Cable EU Spare Part Plug Power Cable UK Spare Part Plug Screw for Rub...

Page 14: ...and maintenance this product is warranted for 24 months against defective materials or workmanship The 24 month warranty period begins from the delivery date of this product For product quality or per...

Page 15: ...478 it fishersci com Finland 358 0 9 8027 6280 fishersci fi France 33 0 388 67 14 14 fishersci fr North America Canada 1 800 234 7437 fishersci ca Netherlands 31 0 20 4887 70 00 nl fishersci com Norw...

Reviews: