6
Please use rubber strips to fasten containers onto
the tray when using this instrument to avoid
accidents.
Symmetrical placement
Asymmetrical placement
Cornered placement
Sided placement
Other placement
2.4
Load
Relationships between Speed and
Tilt Angle 0°
Warning:
The actual maximum rotation speed of the instrument
may be affected if the cell culture flask is placed at the
corner or one side of the tray. If such placement is
needed, please make sure to gradually increase the
speed.
Tilt Angle 12
°
Load Weight
(kg)
Placement
of Load
Tilt Angle
Centered
Symmetrical
9.8
9.8
120
120
0
°
0
°
Rotation Speed
(rpm)
8
0
8
0
Load Weight
(kg)
Placement
of Load
Tilt Angle
Centered
Centered
Centered
Symmetrical
Symmetrical
Symmetrical
9.8
9.8
5
5
1
1
100
100
120
120
12
°
12
°
12
°
12
°
12
°
12
°
Rotation Speed
(rpm)
Position of Loads
Place the loads in the recommended positions below:
1. Place load at the center of the tray
2. Place loads symmetrically around the center of the tray
3. Make sure to use rubber strips to fasten containers onto the
tray
Centered placement