For your safety
4
•
Do not centrifuge toxic or radioactive substances or
pathogenic microorganisms without suitable safety
systems.
If microbiological samples of risk group II (according
to "Laboratory Bio-safety Manual" of WHO) are be-
ing centrifuged, aerosol-tight bio-seals have to be
used.
For materials with a higher risk group, more than
one precaution is required.
•
Should toxins or pathogenic substances enter the
centrifuge or its parts, you must perform appropriate
procedures for disinfection (see "Maintenance and
care – Disinfection").
•
Strongly corrosive substances that may cause
damage to materials and reduce the mechanical
strength of the rotor, may be centrifuged only inside
protective tubes.
Handling the centrifuge
•
Use only original accessories for the centrifuge. The
only exception are common glass or plastic centri-
fuge tubes, if they are approved for the rotor speed
and RCF values.
•
Never use the centrifuge unless the rotor is properly
installed.
•
You may use the centrifuge only with a properly
loaded rotor. You must not overload the rotor.
•
Strictly follow the rules and regulations for cleaning
and disinfection
•
If the rotor or the rotor lid shows signs of corrosion
or
wear, you must stop using it.
•
Never open the lid manually if the rotor still turns.
•
You may use the emergency lid release only in
case of emergency, e.g. during an interruption of
power supply
(see chapter "Troubleshooting").
•
Never use the centrifuge with the lid open.
•
Never use the centrifuge if the front panel has been
partially or totally removed.
•
Changes in mechanical or electrical components of
the centrifuge may be carried out only by individuals
authorized by Fisher Scientific
.
Summary of Contents for accuSpin 3
Page 1: ...accuSpin 3 3R OPERATING INSTRUCTIONS...
Page 10: ...For your safety 6 Notes...
Page 16: ...Description of the accuSpin 12 Notes...
Page 20: ...Rotors and accessories 16...
Page 32: ...Rotors and accessories 28 Notes...
Page 38: ...Operation 34...
Page 68: ...64 Notes...
Page 72: ...68 Acceleration profiles Micro liter Rotor 7500 3337 speed rpm time s...
Page 73: ...69 Deceleration profiles Micro liter Rotor 7500 3337 speed rpm time s 9 1...
Page 76: ...72...