Maintenance and care
46
During cleaning liquids and espe-
cially organic solvents should not
come into contact with the drive
shaft and the ball bearing.
Organic solvents may decompose
the lubricant of the motor bearing.
The drive shaft may block.
Instruments with refrigeration unit:
If a strong ice sheet is present in the internal
chamber, be sure to remove all condensate
after defrosting!
Please control and clean the venting slots
regularly!
Before cleaning the venting slots
please disconnect the centrifuge
from the mains supply.
Please pull mains plug!
Disinfection
If a centrifuge tube containing infectious material leaks
during a run, you have to disinfect the centrifuge im-
mediately.
Infectious material could enter the centri-
fuge if spills or tube breakage occur
.
Danger of infection may occur upon
contact
!
Take appropriate protective meas-
ures for personnel!
Mind the permissible filling volumes and
loading limits for the tubes
!
In case of contamination the operator has
to make sure, that no further persons are
jeopardized
!
Contaminated parts have to be decontami-
nated immediately
.
If required further protective measures
have to be initiated
.
Summary of Contents for accuSpin 3
Page 1: ...accuSpin 3 3R OPERATING INSTRUCTIONS...
Page 10: ...For your safety 6 Notes...
Page 16: ...Description of the accuSpin 12 Notes...
Page 20: ...Rotors and accessories 16...
Page 32: ...Rotors and accessories 28 Notes...
Page 38: ...Operation 34...
Page 68: ...64 Notes...
Page 72: ...68 Acceleration profiles Micro liter Rotor 7500 3337 speed rpm time s...
Page 73: ...69 Deceleration profiles Micro liter Rotor 7500 3337 speed rpm time s 9 1...
Page 76: ...72...