Disassembly of Instrument Parts
The index numbers stated in ( ) reappear within the breakdown drawings and the spare part lists.
5.1
Dismantling the Housing
5.1.1 Front panel (201
•
Open the door and pull out the mains plug
•
Remove the deposit (105)
•
Unscrew the two Philips screws from the top and bottom of the front panel
•
Grip the front panel's upper frame on both sides, pull forwards and depose it face down in front
of unit
•
Pull out plug
XI
and loosen the screen of the key and indication board
•
Refitting the front panel is performed analogously in reverse order
5.1.2 Dismantling the Back Panel (101)
1. 120V models
•
Open the door and pull out the mains plug
•
Unscrew the cross-recessed screws (7x) from the rear panel
•
Take off the rear panel and unplug the cable connection to the main board at mains terminal
L2
and
N
•
Unscrew the ground connection and put down the rear panel
•
Refitting the rear panel is performed analogously in reverse order
2. 230V models
•
Open the door and pull out the mains plug
•
Unscrew the cross-recessed screws (7x) from the back panel
•
Take off the back panel and pull the cable inlet and the mains cable downwards of the slot
•
Unplug the ground connection and put down the back panel
•
Refitting the back panel is performed analogously in reverse order
5.1.3 Dismantling the Lid (110)
•
Open the door, pull out the mains plug and remove the rear panel (see 5.1.2)
•
Unscrew 4 screws (114) from both hinges and remove screws and washers
•
Push both bolts (116) carefully through the gas lid stay holder (117) from out to inner side
•
Remove the door
ATTENTION!
For door replacement transfer both hinges (115), the locking
bolt (120) and handle (123) onto the new lid. Screw will be reached after removing the blue
handle part (123)
•
Refitting the lid plate is performed analogously in reverse order
5.1.4 Dismantling the Housing (220)
•
Open the door, pull out the mains plug and remove the front panel (see 5.1.1)
•
Remove the back panel (see 5.1.2)
•
Remove the door (see 5.1.3)
•
Unscrew the lid lock device and depose it carefully
•
Remove all fastening screws from the casing
•
Remove all leads from the grounding connectors inside the front casing
•
Lift the front casing upwards via the gas lid stay, at the same time pull the housing a little bit
sideways
•