3.5.3
Suction Pressure Test
•
Dismantle the front panel (see section 5, page 1)
•
Screw pressure gauge onto measuring nozzle (suction side on compressor) inflation
pressure (pressure balance after longer standstill) approx. 3 bar at ambient temperature
•
Set temperature to 0°C and start
•
Reading of suction pressure after 2 min. is approx. 0.5 - 0.8 bar
3.5.4
Component Parts of Unit (listed in flow direction)
•
compressor with Schrader valve (low-pressure side)
•
condenser with radial fan
•
dryer cartridge
•
capillary tube (not interchangeable)
•
evaporator integrated in the spin chamber
•
suction line