accuSpin
1R
Calibration Certification
accuSpin
1R
Page 1
Preventive Maintenance Checklist Complete
Repair and Calibration Recertification
Account Name:
Account Address:
Instrument Model:
Serial Number:
Report Number:
Date:
Performance Certification
SPEED
Standard Test
Rotor Used:
Set Speed (rpm)
Tolerance
Displayed
Speed
(rpm)
Measured
Speed
(rpm)
Meets Spec.
(y/n)
Passed
(Initials)
3000
+/-5%
15000
+/-5%
Custom Test
Rotor Used:
+/-5%
+/-5%
TEMPERATURE
Standard Test
Rotor Used:
Calibration
Temp.
(
°
C)
Tolerance
(K)
Displayed
Temp.
(
°
C)
Measured
Temp.
(
°
C)
Meets
Spec.
(y/n)
Passed
(Initials)
+2
+/- 2
+20
+/- 2
Custom Test
Rotor Used:
+/- 2
+/- 2
+/- 2
TIME
Standard Test
Rotor Used:
Timer Set
(min.)
Tolerance
(sec.)
Measured
Time
(m/s)
Meets
Spec.
(y/n)
Passed
(Initials)
10
+/- 30
Custom Test
Rotor Used:
+/- 30
+/- 30
+/- 30
TEST EQUIPMENT
Type
Model
Serial Number
Date of Next Calibration
Stopwatch
Temperature Meter
Photo Tachometer
Safety Tester
SAFETY TEST
Insulation resistance
>10 Megaohm
Earth Conductor Resistance
<0.2 Ohm
Accessible Current
<3500µA
Revised 15/04/03
Certified By:_____________________________ Reviewed By:________________________________ Expiration Date ______
Signature Date Customer Signature Date
Page 1 of 1
Use ONLY Black permanent ink when filling out.