FISCHER MS13 Operating Manual Download Page 17

FISCHER Mess- und Regltechnik GmbH

Commissioning | 3

BA_EN_MS13

17 / 32

NOTICE

Adjustment of the output signal

The basic measuring range (indicated on the type label) and the type of output
signal (voltage / current) are not variable.

The parameters 

MB start

 and 

MB end

 initially define the two pressures

between which the output signal will change at all. Both values are adjustable
across the entire basic measuring range. The set values also refer to the pres-
sure in the respective unit. However, the signal values (current / voltage) for
‘Start of measuring range’ and ‘End of measuring range’ are fixed.
If 

MB start

 is smaller than 

MB end

, this is called an increasing characteristic

curve; the output signal increases as the pressure increases.
If 

MB end

 is smaller than 

MB start

, this is a decreasing characteristic curve and

the output signal decreases as the pressure increases.
The difference between the values 

MB start

 and 

MB end

 must be at least 25 %

of the basic measuring range.
You can select a unit other than the unit of the basic measuring range with the
parameter 

Unit

. The user should remember however that not every unit is suit-

able. The conversion is automatic.
The parameter 

Limit

 allows the display, output and switching points to be lim-

ited to the range between Start of measuring range and End of measuring
range. This makes sense when content is measured to avoid "negative con-
tents". If Limit is set to "no", those measured values that are greater or smaller
than the end values are shown.

3.3.4 Menu Level Output

Parameter name

Description

Value range

min. output

min. output

0.0 … 21.0 mA or
0.0 … 11.0 V

max. output

max. output

Error signal

Measuring range unit

The parameters 

min. output

max. output

 and 

error signal

 define the limits of

the output signal that may not be undercut or exceeded regardless of the pres-
sure. The limit values take priority over the range defined by the 

MB start

 and

MB end

 parameters! These parameters primarily serve to prevent error mes-

sages in downstream systems caused by brief overstepping of measuring
ranges.
The parameter 

min. output

 is usually only used for devices with an output sig-

nal 4…20 mA because frequently values of below 3.8 mA are evaluated as er-
ror signals.
The 

max. output

 value can be used for the voltage and current to limit the max-

imum value.
The value defined via the parameter 

Error signal

 is issued if the device detects

an internal error and can no longer work correctly. It should be noted here that
not all potential errors and faults can be detected by the device itself.

Summary of Contents for MS13

Page 1: ...Operating manual MS13 Digital pressure transmitter switch with colour change display 09005885 BA_EN_MS13 Rev ST4 A 10 16 09005885...

Page 2: ...sing electronic sys tems or any other form print photocopy microfilm or another process without the written consent of the company Fischer Mess und Regeltechnik GmbH Bad Salzuflen Reproduction for int...

Page 3: ...mmissioning 12 3 1 General 12 3 2 Control elements 12 3 2 1 LC display 12 3 2 2 Keyboard 13 3 3 Menu levels 14 3 3 1 Menu Level Switch points 14 3 3 2 Menu Level Input 15 3 3 3 Menu Level Measuring 16...

Page 4: ...lication areas Technical facility equipment TGA Process engineering Process technology Environmental technology 1 3 Intended use The MS13 is a pressure transmitter switch for measuring relative pressu...

Page 5: ...6 7 Code U Process connection Illustration 1 Product Overview 1 Colour change display 2 Bayonet ring housing 3 Keyboard 4 Process connection 5 Blowout plug 6 M12 plug 1 7 M12 plug 2 Process connection...

Page 6: ...of the information that is stated The data shown is purely fictive but does correspond to the actual conditions For more information please see the order code at the end of these instructions MESS UN...

Page 7: ...nder pressure The pressure deforms the measuring mem brane causing a change in resistance on the attached measuring bridge This change is evaluated by the device s electronics and transformed into the...

Page 8: ...llation however any installation posi tion is possible To guarantee safe working conditions during installation and maintenance suit able stop valves must be fitted in the system see accessories By me...

Page 9: ...pe G ca 180 20 56 155 ca 200 G G ca 275 ca 130 20 56 G SW27 Manometer connection Connection spigot according to DIN EN 837 1 Illustration 7 Siphon MZ1 The transmitter must be positioned below the meas...

Page 10: ...b Settable damping reactor In operating mode the damping throttle must be set so that the output signal follows the pressure changes with a delay G G 20 66 19 SW27 Setting screw Connection spigot acco...

Page 11: ...not connect the connector if strained 3 Leiterschaltung Input Output AC DC Sensor Signal Ub Ub Sig Sig Ub Ub p 1 3 2 4 Connector 1 Supply and output signal Illustration 10 M12 connector 5 pin Item Des...

Page 12: ...the pressure connections do not leak before commissioning 3 2 Control elements 3 2 1 LC display 1 3 4 2 Illustration 12 LC display 1 Measurement display 4 6 digits 2 Bar chart display 3 Status display...

Page 13: ...e value is saved with the button To leave a menu level or the entire menu select the parameter Quit and press Example Switch on point set switchpoint 1 In normal mode press the button to enter the men...

Page 14: ...ystemyou can change to the respective na tional language using the language parameter You can see which languages are supported there 3 3 1 Menu Level Switch points Parameter name Description Value ra...

Page 15: ...hysteresis function If SP1 On SP1 Off the output switches on if the measured value exceeds SP1 On and off if the measured value undercuts the same value SP1 Off If SP1 On SP1 Off the output switches o...

Page 16: ...lue x 2x x 2x Zero point window Illustration 16 Zero point window It makes sense to set the Offset zero point displacement if without differential pressure remove measuring line the display shows a va...

Page 17: ...arameter Limit allows the display output and switching points to be lim ited to the range between Start of measuring range and End of measuring range This makes sense when content is measured to avoid...

Page 18: ...Value range Function Function Value square rooted MB decimal pl Measuring range decimal places 1234 123 4 12 34 1 234 12345 123456 MB start Measuring range start 9999 9999 MB end Measuring range end 9...

Page 19: ...ints are used in the table A table is made up of at least 3 max 30 support points NOTICE Number of value pairs If the number of value pairs is changed the table is initialised again and the ex isting...

Page 20: ...e set time is only valid if the parameter Colour is not set to off Amongst other things the legibility of the display depends on the temperature and the reading angle To ensure optimised legibility th...

Page 21: ...alid if the parameter Colour is not set to off The lighting can be switched on permanently with the value 0s The legibility of the display can be adjusted using the parameter Contrast When the contras...

Page 22: ...tch Colour change yellow to red ME MB end Measuring range end In this menu the same parameters are used as those described in the previous sections NOTICE Unused range If a range is not to be used the...

Page 23: ...y The operating time is shown in days d and hours h A Password can be used to protect the menu against unauthorised access The password is a figure from 1 to 999 The input 0 means that no password is...

Page 24: ...5 FS 1 0 FS 0 10 bar 40 bar 70 bar 0 5 FS 1 0 FS 0 16 bar 40 bar 70 bar 0 5 FS 1 0 FS 0 25 bar 100 bar 150 bar 1 0 FS 0 40 bar 100 bar 150 bar 1 0 FS 0 60 bar 200 bar 250 bar 1 0 FS Vacuum and measur...

Page 25: ...linearity and hysteresis 4 5 Auxiliary energy Rated Voltage 24 V AC DC Admissible operating voltage Ub 12 32 V DC Ub 16 5 32 V AC Electrical connection 5 pin M12 circular plug connector 4 6 Applicatio...

Page 26: ...Output signal User definable within the basic measuring range 1 Zero point stabilising 0 of the basic measuring range 2 Zero point correction off set of the basic measuring range 3 Implementation of c...

Page 27: ...S1 13 20 7 5 3 2 6 5 2 17 5 9 5 18 NPT 7 16 UNF G G G Illustration 23 Process connection Port Material G B Connection shanks with external thread 1 4404 G B Connection shanks with external thread 1 44...

Page 28: ...1 PSI 09 0 25 bar 0 2500 kPa 0 362 6 PSI 10 0 40 bar 0 4000 kPa 0 580 1 PSI 11 0 60 bar 0 6000 kPa 0 870 2 PSI 31 1 0 bar 100 0 kPa 14 50 0 PSI 32 1 0 6 bar 100 60 kPa 14 50 8 702 PSI 33 1 1 5 bar 100...

Page 29: ...em Design V FKM Fluororubber Viton 5 1 Accessories Order no Designation No of Poles Length 06401993 Connection cable for switching outputs with M12 connector 4 pin 2 m 06401994 Connection cable for sw...

Page 30: ...6 Attachments FISCHER Mess und Regltechnik GmbH 30 32 BA_EN_MS13 6 Attachments Illustration 24 CE_DE_MS13...

Page 31: ...FISCHER Mess und Regltechnik GmbH Attachments 6 BA_EN_MS13 31 32...

Page 32: ...6 Attachments FISCHER Mess und Regltechnik GmbH 32 32 BA_EN_MS13...

Reviews: