30
MMS
®
Inspection SPG High
Calibration
To calibrate with 2 calibration shims, repeat the entire foil calibration step
(steps 1 to 4) with the second foil.
Display description - Foil calibration step (calibration shim)
1 Name of calibration
2 Schematic illustration of the current
calibration step
3 Current calibration step
4 Progress display of the calibration
steps (example of 3 calibration steps,
calibration step 2 current)
5 Currently measured reading
(example)
6 Mean value of the previously
measured measured readings
(example)
7 Deletes the last measured reading
8 Opens the window to set the calibration shim nominal value
9 Battery indicator
10 Set nominal value of the calibration shim used (example)
11 Number of measurements (example)
12 To next calibration step/exit calibration
4
Take the calibration shim from the glass plate
Return the calibration shim (1/2) to the protective sleeve.
Glass plate
Calibration shim
Calibration 001
Foil 1
x
OK
1
2
3
4
5
6
7
8
9
10
11
12
x
77.4
94.29
Foil
n
98
3
SET