MMS
®
Inspection SPG High
29
Calibration
2
Perform 5 to 10 measurements with the calibration shim.
3
Check whether the nominal value of the calibration shim used is shown in
the
Foil
table cell.
►
How to set the nominal value of the calibration shim:
SET >
set nominal
value via
> OK
Calibration 001
78.6
x
OK
Foil 1
A
Placing the gage
Place the gage on the sur-
face. The probe tip must be
touching the glass plate
within the hole.
B
Measurement acquisi-
tion
The measurement acquisition
is reported by a beep and
also by the illumination of the
signal lamp. The new mea-
sured reading appears in the
display.
C
Lifting the gage
Lift the gage from the sur-
face.
Repeat steps A to C for
the next measurement on the uncoated specimen
102.7
x
98.0
n
2
Foil
The display screens are to be understood only as examples
Calibration 001
77.4
x
OK
Foil 1
94.29
x
98.0
n
3
Foil
Calibration 001
77.4
x
OK
Foil 1
94.29
x
98.0
n
3
Foil
Glass plate
Calibration shim