26
MMS
®
Inspection SPG High
Calibration
7.2.3
Zero calibration step
Measurements on the glass plate from scope of supply.
Required material
•
Glass plate 605-306 from scope of supply
Procedure – Zero calibration step
1
Perform 5 to 10 measurements on the glass plate.
Calibration 001
-3.3
x
OK
Zero
A
Placing the gage
Place the gage on the sur-
face.
B
Measurement acquisi-
tion
The measurement acquisition
is reported by a beep and
also by the illumination of the
signal lamp. The new mea-
sured reading appears in the
display.
C
Lifting the gage
Lift the gage from the sur-
face.
Repeat steps A to C for
the next measurement on the uncoated specimen
-3.50
x
n
2
The display screens are to be understood only as examples
Calibration 001
-3.8
x
OK
Zero
-3.58
x
n
3
Calibration 001
-3.8
x
OK
Zero
-3.58
x
n
3
Glass plate