28
MMS
®
Inspection SPG High
Calibration
7.2.4
Foil calibration step (calibration shim)
Measurements on the calibration shim (foil) located on the glass plate directly.
Required material
•
Calibration shim(s) from scope of supply. The hole in the shim marks the
specified measuring area.
•
Glass plate from scope of supply
Procedure – Foil calibration step (calibration shim)
1
Position the calibration shim
Place the calibration shim (Foil 1/2) on the glass plate.
There must only be
one
calibration shim on the glass plate at any one time!
Glass plate
Calibration shim
Hole in the shim