Brief Guide FMP10 / FMP20
Page 15
Adjusting Instrument and Probe
Place the calibration standard 1 (in the ex-
ample with a thickness of 23.7 µm) on the
uncoated specimen and perform measure-
ments.
Make individual measurements at various
points of the reference area. Displayed is
the mean value of all measurements per-
formed for this step.
∧
or
∨
[Entry:
∧∨
]: Use the arrow keys to set the
nominal thickness value of the calibration
standard. The rated value is printed onto
the calibration standard. The rated value
can be set faster if a measurement is per-
formed on the calibration standard and
then the rated value is corrected using the
arrow keys.
This step is not required if - as shown in
the example - the rated value for the thick-
ness of the calibration standard coincides
with the stored thickness.
ENTER
If a corrective calibration is desired with 2
standards, proceed in the same manner
with calibration standard #2.
Otherwise: Use ENTER to end the correc-
tive calibration.
The new characteristic will be computed
automatically and stored.
The instrument is again ready to make
measurements.
ENTER
A confirmation indicating that the correc-
tive calibration has been carried out suc-
cessfully appears.
Pressing ENTER confirms the message.
Key sequ. /
Action
Detail of the LCD
display