Brief Guide FMP10 / FMP20

Page 15

Adjusting Instrument and Probe

Place the calibration standard 1 (in the ex-
ample with a thickness of 23.7 µm) on the 
uncoated specimen and perform measure-
ments.
Make individual measurements at various 
points of the reference area. Displayed is 
the mean value of all measurements per-
formed for this step.

 or 

[Entry: 

∧∨

]: Use the arrow keys to set the 

nominal thickness value of the calibration 
standard. The rated value is printed onto 
the calibration standard. The rated value 
can be set faster if a measurement is per-
formed on the calibration standard and 
then the rated value is corrected using the 
arrow keys. 
This step is not required if - as shown in 
the example - the rated value for the thick-
ness of the calibration standard coincides 
with the stored thickness.

ENTER

If a corrective calibration is desired with 2 
standards, proceed in the same manner 
with calibration standard #2.
Otherwise: Use ENTER to end the correc-
tive calibration.
The new characteristic will be computed 
automatically and stored.
The instrument is again ready to make 
measurements.

ENTER

A confirmation indicating that the correc-
tive calibration has been carried out suc-
cessfully appears.
Pressing ENTER confirms the message.

Key sequ. / 
Action

Detail of the LCD 
display

Explanation

Summary of Contents for DELTASCOPE FMP10

Page 1: ...DELTASCOPE FMP10 ISOSCOPE FMP10 DUALSCOPE FMP20 Guidelines Coating Thickness Material Testing Microhardness Material Analysis...

Page 2: ......

Page 3: ...Install the batteries rechargeable batteries 7 6 2 Probe connection 8 6 3 Instrument On Off 9 7 Adjusting Instrument and Probe 11 7 1 Assigning a New Probe 11 7 2 Normalization 12 7 3 Corrective Calib...

Page 4: ...tensive oper ator s manual can be found on the CD ROM that is supplied with the instrument The following symbols are used in this brief guide Indicates safety information referring to danger for per s...

Page 5: ...non conducting non magnetic coatings on non ferro magnetic electrically conducting base materials Paint lacquer or plastic coatings on for example aluminum copper zinc etc as well as anodized coating...

Page 6: ...obe connector socket Page 8 2 LCD Display Page 4 3 Keys for directly retrieving functions Page 5 4 USB port for connecting a PC 5 ON OFF key to turn the in strument on or off Page 9 6 Cover additional...

Page 7: ...rement uses the eddy current method Padlock Restricted operating mode is enabled i e the keys ZERO and CAL are not active the service functions cannot be retrieved Arrow circle Free running display is...

Page 8: ...rieves the corrective calibration z Changes the information shown on the LCD display z The display will change faster if is pressed for longer than 3 seconds z Changes the information shown on the LCD...

Page 9: ...easuring beginning on Page 16 z Always hold the probe at its grip sleeve right figure z Always place the probe gently and at a right angle on the specimen surface z Slide the grip sleeve to the specim...

Page 10: ...AM3 MN1500 batteries or 4 individual rechargeable batteries 1 2 V 2400 mAh Type AA Using other batteries may lead to instrument damage Use only non damaged batteries rechargeable batteries If the bat...

Page 11: ...ON OFF key on the right side of the control panel The LCD display is not backlit and no characters are visi ble Protect the instrument and accessories from electrostatic charges Electrical discharges...

Page 12: ...s of the LCD display will appear briefly at the same time Page 4 At the end of the monitoring routine the instrument is ready to make measurements The icon of the measure ment method of the connected...

Page 13: ...e connected NF Fe NC NF Dual probe connected and dual method both set i e in the open application both measurement methods can be used to make measurements NF Fe NC NF Dual probe connected and dual me...

Page 14: ...006 Probe changed The probe must be assigned Chapter 7 1 7 1 Assigning a New Probe Key sequ Action Detail of the LCD display Explanation This warning appears briefly after power up if a probe other th...

Page 15: ...n is per formed s Standard deviation n Number of measurements Base material Fe NF Measurements should be performed on an uncoated spec imen Fe or NF display for the type of base material When using du...

Page 16: ...shed for the calibration curve and stored in the instrument Required materials z Reference part uncoated part from the production z Calibration Standards Selecting the Calibration Standards for the Co...

Page 17: ...f base material will be displayed only after the 1st measurement is completed Perform measurements on the uncoated specimen base material normalization Make measurements at various points of the refer...

Page 18: ...er formed on the calibration standard and then the rated value is corrected using the arrow keys This step is not required if as shown in the example the rated value for the thick ness of the calibrat...

Page 19: ...ration respectively 8 Measuring To make a measurement place the probe at a right angle on the specimen surface The probe can be lifted off after the measurement acquisition i e after the read ing appe...

Page 20: ...ument on After pressing ENTER 5 times the identifi cation number 157 appears on the LCD display 2 x Enter the identification number 159 by pressing the arrow key 2 times Use ENTER to confirm the setti...

Page 21: ...ee running mode Instrument mode Restr Mode Analog display Measurement Audible signal Measurement effect External start Measuring mode Standard Area measurement Automatic measurement Display dual metho...

Page 22: ...29 34 51 japan helmutfischer com www helmutfischer com FRANCE FISCHER INSTRUMENTATION ELECTRONIQUE Parc d Activit s Nord du Pas du Lac 5 rue Micha l Faraday F 78180 Montigny le Bretonneux BP 289 F 78...

Page 23: ......

Reviews: