FISCHER DE44 Operating Manual Download Page 7

 

The  table  is  correct  if  the  following  applies  for  all 
signal values: The value is larger than the previous 
value.  Either  larger  (rising  characteristic  curve)  or 
smaller  (falling  characteristic  curve)  apply  to  the 
pressure values accordingly. No transition from ris-
ing  to  falling  characteristic  curves  or  vice  versa  is 
allowed. 

5.3.10  Switch points 

The two switch outputs 

 

 are configured by four 

parameters respectively.  

The  function  of  the  switching  output 

  is  deter-

mined by the parameters 

R1A

R1E

R1D

 and 

R1F

.  

The  function  of  the  switching  output 

  is  deter-

mined by the parameters 

R2A

R2E

R2D

 and 

R2F.

 

R1A

  defines  the  switch-off  point, 

R1E

  defines  the 

switch-on  point  of  switch  output  1.  The  values  are 
set in the valid measuring unit (shown on the right). 

Together, the two parameters R1A and R1E  deter-
mine the switch function of switch output 1:  

If 

R1A

  is  smaller  than 

R1E

,  the  output  switches  on,  if 

the measured value exceeds 

R1E

. It is only switched 

off again if the measured value 

R1A 

is undercut (hys-

teresis function). 

If 

R1A

  = 

R1E

,  the  output  switches  on  if  the  measured 

value exceeds 

R1E 

and off if the measured value un-

dercuts 

R1A

.  

If 

R1A

 is larger than

 R1E, 

the output switches on, if 

R1E 

< Measured value < 

R1A 

applies (window function). 

Both parameters can be set independently over the 
entire range. 

If the measuring unit is switched over, the switching 
points  are  converted  accordingly.  Rounding  errors 
may cause deviations in the last position. 

R1D 

allows  the  reaction  of  the  switch  output  1  to  be 

delayed  by  between  0.0  and  100.0  s.  This  value 
applies equally for switching on and off. 

R1F

  reverse  the  function  of  the  switch  output.  If  the 

value  =  1,  the  switch  output  works  as  an  NO  con-
tact, if the value = 2, the switch out works as a NC 
contact. 

5.3.11  Password 

The last menu item 

-P-

 is used to enter a password. 

A  value  between  001  and  999  can  be  selected  for 
the password. The value 000 cancels the password 
function. 

If a password has been issued, the text 

PAS

 appears 

after 

ESc

  and 

û,

  and  you  need  to  enter  the  correct 

value  using 

û

  and 

þ

,

ÿ

.  You  will  only  arrive  at  all 

other menu items after doing this. In the event of an 
error,  the  display  will  jump  back  to  the  start  of  the 
menu 

ESc

If  the  password  is  forgotten,  it  can  only 
be  reset  by  the  manufacturer  or  over-
written via the PC adapter. 

5.3.12  Display options 

The  parameter 

D0

  enables  the  reading  to  settle  if 

the  measuring  value  fluctuates  heavily.  This  filter 
function  is  similar  to  the 

dAM

  function,  but  only  im-

pacts on the reading not on the output signal. At 

D0

 

= -1 only the switchpoint LEDs are controlled. At 

D0

 

= -2 these are switched off. 

5.3.13  Reset to default 

The  function 

res

  allows  all  settings  to  be  reset  to 

the default settings. The default values can only be 
defined via a PC interface. 

5.3.14  Free unit 

If  the  device  is  designed  for  a  "free"  third  unit 
(membrane  symbol: 

),  the  display  can  be  scaled 

infinitely using the parameters 

MAF

MEF

 and 

dPF

The measuring range defined by the parameters 

MA

 

and 

ME

 is converted to 

MAF

 and 

MEF

. This also takes 

into account the table function ( 

F

 ). The value of 

dPF

 

determines the position of a decimal point. 

5.4 

Parameter overview 

After  switching  on  the  device,  it  will  briefly  indicate 
the  software  version  number  and  then  enters  the 
normal operating mode. By using the middle 

û

 key 

on  the  membrane  keypad  you  can  access  the  pa-
rameter menu. The reading now shows the text 

ESc

By  using  the  right 

þ

  key,  you  can  choose  the  pa-

rameters from the following list one by one:  

Note:  
Depending on the device model, individ-
ual  parameters  may  not  be  available  if 
the device does not have this feature.  

PAS 

Enter password 

(only appears if the password is ac-
tive), value range 000...999 
000 = deactivated 

DAM 

Attenuation 

(Jump response time T90), 
value range 0.0..100.0s 

d0 

Display attenuation 

Value range -2...0...100.  
-2 = Display off,  LED Switch point off 
-1 = Display off,  LED Switch point on 
 0 = Display on,  LED Switch point on 
1...100 Display attenuation 

R1A 

Switch-off point 

From switching output 

 

 

 

Summary of Contents for DE44

Page 1: ...zed personnel familiar with the in stallation commissioning and operation of this product Specialized personnel are persons who can assess the work they have been assigned and recognize potential dang...

Page 2: ...eed between the manufacturer and user 3 Description of the product and func tional description 3 1 Function diagram 3 2 Design and mode of operation The device is based on two piezo resistive sensor e...

Page 3: ...ilters need to be installed at suitable points The pressure sensing lines need to be kept as short as possible and installed without sharp bends to avoid interfering delay times If the pressure sensin...

Page 4: ...tions step by step Depending on the device model 1 some menu items may not be available For example all characteristic curve functions are faded out in the menu if the de vice does not have a signal o...

Page 5: ...isplay output signal and switch points but not on the measuring cell it self This parameter can be used to set the reaction time to the pressure jumps The value range is 0 0 s to 100 0 s But with maxi...

Page 6: ...means of hydrostatic pressure measure ments are two typical examples of this In these cases you might want to change the output signal of the transmitter to a non linear characteristic curve so that t...

Page 7: ...ue 000 cancels the password function If a password has been issued the text PAS appears after ESc and and you need to enter the correct value using and You will only arrive at all other menu items aft...

Page 8: ...e of the de vice can be shown sensibly MA Start of measuring range The measuring value is set in that the output signal is minimal e g 0V 0mA or 4mA ME End of measuring range The measuring value is se...

Page 9: ...must not be exposed to mechanical shocks It shall be transported only in packaging specifically intended for transport 8 Service All damaged or faulty devices shall be directly sent to our repair depa...

Page 10: ...4 20 mA 0 10 V DC Admissible apparent ohmic resistance RL Ub 4 V 0 02 A Ub 26V RL 1100 Ub 26V RL 2 K Ub 15V RL 10 K Ub 12 15V Characteristic curve can be programmed Power consumption approx 2 W VA Di...

Page 11: ...ith 3 30 support points Password 001 999 000 no password protection Comments 0 For values from 0 2 the display switches between dP1 and dP2 rhythmically 1 Measuring values around zero are set to zero...

Page 12: ...5 6 0 40 mbar 5 7 0 60 mbar 5 8 0 100 mbar 5 9 2 5 2 5 mbar A 6 4 4 mbar A 7 6 6 mbar A 8 10 10 mbar A 9 16 16 mbar B 1 25 25 mbar B 2 40 40 mbar C 5 60 60 mbar B 3 100 100 mbar B 4 0 400 Pa D 7 0 60...

Page 13: ...40 mbar 5 7 0 60 mbar 5 8 0 100 mbar 5 9 2 5 2 5 mbar A 6 4 4 mbar A 7 6 6 mbar A 8 10 10 mbar A 9 16 16 mbar B 1 25 25 mbar B 2 40 40 mbar C 5 60 60 mbar B 3 100 100 mbar B 4 0 400 Pa D 7 0 600 Pa D...

Page 14: ...sured value display contact elements 3 1 2 digit LED 2 relay contacts 3 3 1 2 digit LED 2 semiconductor switches 6 Electrical connection M12 plug connection M Assembly option Standard attachment boreh...

Page 15: ...14 Declaration of Conformity...

Page 16: ...ject to change without notice Changements techniques sous r serve Fischer Mess und Regeltechnik GmbH Bielefelder Str 37a D 32107 Bad Salzuflen Tel 49 5222 9740 Fax 49 5222 7170 eMail info fischermesst...

Reviews: