20
MMS
®
Inspection DPM High
Measurement
6.1 Before you start
•
The gage is switched on (
key
)
•
Temperature probe
•
The gage surface temperature probe is used: The protective cap has
been removed
•
A magnetic surface temperature probe is used (example): The
temperature probe is connected to the gage
•
The required batch is opened ( key in the main menu, see also
page 18)
6.2 Measurement procedure
•
Air temperature (Ta)
, measured continuously
•
Relative humidity (%RH)
, measured continuously.
•
Surface temperature (Ts)
Measuring with the gage surface
temperature probe: Press the surface
probe on the surface until the value for Ts
hardly changes any more.
Alternatively, you can measure the
surface temperature by connecting a
Type K surface probe to the gage
•
To manually save all climate factors
:
>
•
To continuously log all climate factors
(log func-
tion):
>
The top left corner of the window shows the
remaining time in minutes up to the next save
(e.g.: save interval 3600 min (= 6 h), every 3600 minutes, the climate
factors measured and calculated at this point in time are saved;
remaining time to the next save is 3598 minutes and 45 seconds)
•
Stop continuous logging (log function):
Gage
surface tem-
perature
probe
Surface
Spring-loaded temperature sensors
3598:45
Batch 001
%RH
34.9
Ta
26.1
°C
A