Wet Measurement:
The display for measurements using wet analysis are identical as for dry. The
water content in the flue gas will have to be determined and entered for reading accuracy. Refer to
Setup Menu.
3.3 RECALIBRATE
When the Firetron 1000 is first switched on, it automatically performs a zero calibration. A zero cali-
bration may be initiated manually by selecting recalibrate at the main menu. The ABORT function
key allows early termination of the zero cal cycle
(Note: Aborting the recalibration cycle may
cause
incorrect gas readings).
The completion or abortion (defaults to previous calibration) of a
zero calibration cycle records the voltage outputs of the cells as exposed to fresh air. The cell zero
values are displayed under: DIAGNOSTICS -> CELL ZEROS, or may be printed out using the
DIAGNOSTICS -> REPORT facility.
NOTE: Cell zeros are normally 0.0v (offset should be less than ±20mV), except for: O
2
(span
value) which should be around 1.6V TO 2.0V. Refer to section 3.5.2 diagnostic cell zeros.
3.4 SETUP
The Setup menu enables both essential measurement parameters and units to be setup. See display
below.
MENU
WET
NOx
F1
F2
F3
F4
XXX ppm
ENTER: NEXT SCREEN
FUEL EFF
LOSS
EXCESS AIR
XX.X%
XXX %
XX.X%
HELP
ZOOM
LOG
O2NORM ON
CO2
XXX %
MENU
F1
F2
F3
F4
HELP
ABORT
CALIBRATING PROBE
CAL TIME REMAINING: 180 SEC.
MENU
HELP
GASES
SYSTEM
FUEL TYPE
FUEL DATA
DATE AND TIME
PASSWORD
PRINTER
SERIAL PORT
F1
F2
F3
F4
GAINS & CONSTANTS