![Finero DCP-1500-05 User Manual Download Page 17](http://html1.mh-extra.com/html/finero/dcp-1500-05/dcp-1500-05_user-manual_2289582017.webp)
FST-DCP-1500-05 HIGH VOLTAGE POWER SUPPLY
2283
USER MANUAL
22/01/2010 12:52 PM
AT
UserManual FST-DCP-1500-05 HighVoltagePowerSupply 100122EN
ver1.doc
16/27
------------------------------------------------------------------
Finero Oy
–
The Quality Control Company
Tel. +358-5-7522 200
Haltonintie 5, FIN-47400 KAUSALA, FINLAND
Fax +358-5-3638 326
www.qacontrol.com
8 MALFUNCTIONS
8.1 General
If any faults or a malfunction occurs in the operation of the Tester, the following protecting
devices and functions should be checked. Printed circuit board fuses are TE5 types and can
pull up from socket. Check the fuse by a multimeter and replace with the similar one if neces-
sary.
8.2 No indicator is lit
Check the power cord connectors are properly plugged at both ends
Check the voltage in the mains socket.
Check the main fuses (F1 and F2) of the Tester. The fuses are located in the fuse box below
the power cord connector. The box can be opened by removing first the mains cord and then
pulling the fuse box about 10 mm outwards. The fuses will be raised slightly. Check the fuse
by a multimeter and replace with the similar one if necessary.
8.3 Power indicator is lit but the display is not
Check the fuse of the power unit. Fuses are in control board middle of tester. There are left
side two fuses FU1 and FU2.
8.4 Indicators are lit but buttons don't function
Check the safety circuit.
Check set values in “Setting” menu.