Electromagnetic immunity
The Fiagon system is intended for use in the electromagnetic environment specified below. The
customer or the user of the system should assure that it is used in such an environment.
Immunity Test
ESD
IEC 61000-4-2
Burst
IEC 61000-4-4
Surges
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines,
IEC 61000-4-11
Power
frequency
(50/60 Hz)
Magnetic field,
IEC 61000-4-8
IEC 60601-1-2 Test
Level
+/-6kV cd
+/-8kV ad
+/-2kV mains
+/-1kV I/O
+/-1kV dm
+/-2kV cm
Reduction to:
5% for 10 ms/positive
amplitude
5 % U
T
for 10ms/
negative amplitude
5 % U
T
for 10ms/
negative amplitude
40 % U
T
for 100 ms
70 % U
T
for 500 ms
5 % U
T
5000 ms
3 A/m
Compliance Level
Level
+/-6kV cd
+/-8kV ad
+/-2kV mains
+/-1kV I/O
+/-1kV dm
+/-2kV cm
Reduction to:
5% for 10
ms/positive
amplitude
5 % U
T
for 10ms/
negative amplitude
5 % U
T
for 10ms/
negative amplitude
40 % U
T
for 100 ms
70 % U
T
for 500 ms
5 % U
T
5000 ms
3 A/m
Electromagnetic Environment
Guidance
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
U
T
is the a.c. mains voltage prior to application of the test level.