29
Instructions for Use
Copyright ©
2018 Products
DocumentID
Version
Date
29 / 34
Fiagon GmbH
E 01 1103
DOC-18-00147 1
09-2018
EMC compliance statement
The Fiagon Navigation System needs special precautions regarding EMC and needs to be installed and put into
service according to the following described EMC information.
Electromagnetic emission
The Fiagon system is intended for use in the electromagnetic environment specified below. The
customer or the user of the system should assure that it is used in such an environment.
Emission Test
RF-emissions following
IEC/CISPR 11
Compliance
group 1
Electromagnetic Environment Guidance
The system uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF-emissions following
IEC/CISPR 11
class B
The Fiagon system is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
The intended use environment is hospital
environment.
Generation of mains
harmonics
Following IEC 61000-3-3
class A
Generation of voltage
Fluctuations/ flicker following
IEC 61000-3-3
Complies