Modes of operation
The Fiagon Navigation System has three modes of operation that are automatically switched dependent on the
situation and user interaction.
•
Navigation Mode (for navigation interaction)
•
Mouse mode (for software operation interaction)
•
Full screen video mode
Navigation mode
In this mode the navigation information of the instrument is displayed in the image data. The mode is activated
automatically when the instrument is being guided in the measurement field of the navigation sensor.
The status indicator turns green
The display is optimized for displaying the navigation information. All software interaction buttons and the mouse
cursor are blinded out. The navigation instrument has the focus.
Full screen Video mode
This mode is configured in the applications (modules) offered by Fiagon that utilize a video signal of a medical
video device (endoscope or microscope).
In this mode the connected video signal is displayed in full screen on the display. The mode is activated when
no navigation information is available for 3 seconds. That is when the navigation instrument is currently not used
and is guided out of the measurement field of the navigation sensor
The user automatically gets the available video information in optimized full screen display.
Mouse mode
In this mode the user can interact with the functions of navigation software by using the mouse. The mode can
be activated at any time by clicking once on any of the mouse buttons.
The interaction buttons and the mouse cursor are faded in. The mouse cursor has the focus and operation by
mouse is possible.
The mouse mode automatically returns to navigation mode or Full screen video mode after 5 seconds of inactivity
of the mouse.
NOTE:
If you want to operate the system with the mouse (mouse-mode), click once to leave
the full-screen mode or navigation mode and to display the cursor. You can operate the
software now with the mouse. The system will exit this mode automatically 5 seconds after
you stopped using the mouse.
The status indicator is red