14
Instructions for Use
Copyright ©
2018 Products
DocumentID
Version
Date
14 / 34
Fiagon GmbH
E 01 1103
DOC-18-00147 1
09-2018
•
surgical turbines
•
RF surgery
•
DECT phones, cellular phones
•
RFID tags
In close vicinity, (e.g. operating area/ 2 m radius, 3.3 feet)
•
Diathermy devices
•
Security systems such as metal detectors
•
Other image guided surgery systems with EM tracking technology
In adjacent rooms:
•
MR scanner
For EMC technical specifications of the system refer to the related chapter in this document.
NOTE:
RFID tags are not likely to interfere with the system unless they are positioned directly
between the field generating device and the sensor. Make sure the instruments and equipment
used for the surgery do not contain any RFID tags that will be in the measuring field. Consult
the Instructions for use of the devices to check whether and where RFID technology is used.