L011-44
Page 8 of 9
2019-05-13
Rev. E1
5. Remove platform and drill burr hole.
6. Reattach platform.
7. Attach center drive positioner to center hub with tab oriented 45°, 90°, or 135° relative to screw.
Make sure positioner is completely seated in the hub.
8. Attach drive to positioner and secure following drive instructions.
9. Perform MER and implant lead.
10. Remove drive from positioner, then remove platform.
11. Use the standoff wrench to support the standoff while twisting the driver clockwise to secure it
to the standoff. Turn standoff wrench counterclockwise to remove standoff.
12. Use the anchor wrench to support the anchor while twisting the driver clockwise to secure it to
the anchor. Turn anchor wrench counterclockwise to remove anchor.
13. Dispose of used anchors, Standoffs and thumbknobs according to hospital protocol.