Illustrative Procedure

1.  Attach and handtighten standoffs to anchors.

 

 
 
 
 
 
 
 
 

2.  Attach platform to standoffs with thumbknobs.

 

 
 
 
 
 
 
 

3.  Attach the center hub to the platform using 2 screws at 90 degrees apart. Make sure the hub is 

completely seated in the platform. The hub will sit approximately 1/8” above the top surface of 
the platform.

 

 
 
 
 
 
 
 

4.  Use the burr hole entry marker with burr hole bushing and mark the skin and skull, then 

remove.

 

 
 
 
 
 
 
 
 

a

a

b

a

b

c

d

 

 

WARNING:

 Do not 

overtighten thumbknobs. Only 
turn an additional 1/4 turn after 
contacting mating surface. 

b

Summary of Contents for STar 66-IT-VP-01

Page 1: ...Drive Adapter Kit for microTargeting Platforms With Intergrated Ring FHC Inc 1201 Main Street Bowdoin ME 04287 USA Fax 1 207 666 8292 24 hour technical service 1 800 326 2905 US Can 1 207 666 8190 FH...

Page 2: ......

Page 3: ...ev E1 2019 05 13 Table of Contents Indications for use Intended use and Other devices required Symbol Key Inventory Cleaning and Sterilization Illustrative Procedure Track Selection Track Offsetting C...

Page 4: ...tion electrodes or other instruments in the brain or nervous system Intended use STar Drive Platform Adapter Kit is intended for use by a neurosurgeon in an operating room environment to position and...

Page 5: ...fication Probe 66 IT VP 01 Inventory 1 3 2 4 5 6 7 8 9 10 12 A 13 14 15 16 11 Purchased Separately A STar Drive M E shown or STar Drive M A Various or STar Drive sold in kits The products referenced i...

Page 6: ...ding to manufacturer s recommendations Asepti Wash Plus liquid 2 5 ml per liter or 1 4 oz per gal using warm tap water 2 Separate the Drive positioners hubs and Lead Holder and immerse them in the was...

Page 7: ...latform using 2 screws at 90 degrees apart Make sure the hub is completely seated in the platform The hub will sit approximately 1 8 above the top surface of the platform 4 Use the burr hole entry mar...

Page 8: ...form MER and implant lead 10 Remove drive from positioner then remove platform 11 Use the standoff wrench to support the standoff while twisting the driver clockwise to secure it to the standoff Turn...

Page 9: ...using positions A H Examples 3mm Offset Positioner in the G position indicated by the dark blue dots in the offsetting chart 3mm Offset Positioner in the B position indicated by the magenta dots in th...

Reviews: