Automated Cleaning Protocol:
(Requires STar™ SteriSuite)
Platform Adapter Kit Sterilization instructions:
Method
Steam
Protocol
Gravity wrapped:
(in 2 Layers of 1-ply Polypropylene wrap
[1]
)
Exposure Time: 30 minutes at 132°C (270°F)
Minimum Dry Time: 60 minutes
[1] Cycle was validated using Halyard Health H300 wrap
Prevacuum wrapped:
(in 2 Layers of 1-ply Polypropylene wrap
[2]
)
Preconditioning Pulses: 3
Exposure Time: 4 minutes at 132°C (270°F)
Minimum Dry Time: 60 minutes
[2] Cycle was validated using Halyard
Health H400 wrap
Prevacuum wrapped:
(in 2 Layers of 1-ply Polypropylene wrap
[3]
)
Preconditioning Pulses: 3
Exposure Time: 18 minutes at 134°C (273°F)
Minimum Dry Time: 60 minutes
[3] Cycle was validated using Halyard Health H300 wrap
User should be aware that the effects of unvalidated sterilization protocols could result in damage
to the components and affect their functioning or performance. The STar™ Drive Adapter Kit is not
validated for use with alternative sterilization protocols, and FHC does not recommend or endorse
their use. Users with questions regarding this safety issue should contact FHC’s Technical Service
Department at 1-207-666-8190.
Manual Cleaning Protocol:
(Requires STar™ SteriSuite)
1. Prepare the detergent according to manufacturer’s recommendations: Asepti Wash Plus liquid
(2.5 ml per liter or 1/4 oz per gal), using warm tap water.
2. Separate the Drive, positioners, hubs, and Lead Holder and immerse them in the wash solution
for a minmum of 5 minutes. Actuate the devices during the soak.
3. Using a clean soft cloth that has been soaked in the detergent, wipe the tray, and its insert, to
remove any visible soil. Use a soft bristle brush and syringe to reach hard-to-clean areas.
4. Place the Drive and its components back in the tray.
5. Prepare the detergent in a sonication unit according to manufacturer’s recommendations: Asepti
Wash Plus liquid (2.5 ml per liter or 1/4 oz per gal).
6. Immerse the tray in the sonication unit and sonicate for a minimumof 10 minutes.
7. Rinse all components with running reverse osmosis/de-ionized water to remove any residual
detergent.
8. Dry components using a clean soft cloth.
9. Visually inspect to ensure all visible soil is removed.
Cleaning and Sterilization
Phase
Recirulation Time
Water Temperature
Ecolab Inc. Detergent
(2.5 ml/l or 1/4 oz/gal)
Ecolab Inc. Detergent
(2.5 ml/l or 1/4 oz/gal)
Pre-Wash 1
2:00 min
Cold Tap Water (16°C max)
N/A
N/A
Enzyme Wash
2:00 min
Hot Tap Water (43°C min)
Asepti Wash Plus
Sekusept AR
Wash 1
2:00 min
65.5°C (Set Point)
Asepti Wash Plus
Sekusept AR
Rinse 1
2:00 min
Heated Water (66.0°C)
N/A
N/A
Pure Water Rinse
0:10 min
Heated (66.0°C)
Asepti Rinse
Sekusept FNZ or
Sekumatic Multiclean
Dry
7:00 min
115°C
N/A
N/A