User Guide & Operator instructions
P194
Re0
January 2021
C SERVICE INPUTS
Brewer Sensors and Keypad
POSITION Program Items
Factory set Default
Programming Range
Increments
Notes
C1
LLC Probe
continuity
Direct read
Reading of tank
water resistance in
TDS
≈850
- LOW
≈1600
-HIGH
Nominal values
C2
Brew Basket
Sensor
Direct read
YES or NO
C3
Tank
Temperature
Direct read
Hot water tank
temperature
Actual values
C4
USB Drive
NO
C6
Keyboard Test
Calibrate
Checks buttons
under membrane
cover
YES/NO
Follow directions on
the touch screen
.
D SERVICE OUTPUTS
Test Valves and Heaters; Set screen brightness
POSITION Program Items Factory set Default
Programming
Range
Increments
Notes
D1
Heater SSR Test Press button 2 to test
(button 1 stops test)
Activates heater
Default is 10 sec..
/-
OFF or ON
Energizes Heater(s)
WARNING!
Service use only.
D2
Fill Valve Test
Press button 2 to test
(button 1 stops test)
Activates valve
Default is 10 sec.
/-
OFF or ON
Press To Test
D4
Brew Valve Test (Press to test)
Activates valve
Default is 10 sec.
/-
OFF or ON
Runs valve to verify flow.
NOTE: Have container
under brew basket.
D8
Basket Lock
(Press to test)
/-
OFF or ON
Press To Test
Single series displays right side only Left Valve display is only for twin side brewer.
D12
LCD Brightness
Brightness=90%
20-100%
5%
Adjust LCD screen
brightness only-Not for
LEDs under buttons
D13
LED Brightness
Brightness=60%
20-100%
5%
Adjust LED button
brightness only-Not for
the screen display LCD
.
E CALIBRATION
Brewer Sensors and Keypad
POSITION Program Items
Factory set Default
Programming Range Increments Notes
E1
Ready Temp.
Offset
-4°F
-2°C
-2° to -9°F
-1° to -5° C
1°F
1°C
Compensates output to
measured temperature
E2
LLC Sensitivity
NORMAL
(“NORMAL” for most
water)
HIGH
(Biased for reverse
osmosis water or
very pure water)
NORMAL
HIGH
Liquid level control
sensitivity.
High,1300Ω is for reverse
osmosis water or very
pure water.
E3
Slow flow rate
from supply
ON
OFF/ON
/-
YES or NO
Trims fill system for low
supply or Flojet use
E5
Brew valve flow
rate:
1.60L/0.42G
1.30-1.90Liter
0.34-0.52G
0.05L
0.013G
Adjusts flow rate
]
Use this formula to compensate for minor discrepancies in actual volume versus programmed volume.
Default Brew Valve Flow Rate—All CBS-1200 Brewers
CBS-1220
L/min
Range
Right Brew Valve FR 1.60
+1.3 L to 1.9 L
Use the formula above to determine the correct setting
Set lower to increase volume, higher to decrease volume.
See “PROGRAM” E5 For valve settings and calibration. Factory set brew valve flow rates are in liter/min