© Ferno-Washington, Inc. / 234-3622-06 / March 2019
61
7.4 Set the Loading Height
The user can set the iN∫X so it stops automatically at the proper
loading height for the assigned ambulance. To set or change
the loading height:
1. On a level surface, raise or lower the iN∫X until the
loading wheels roll directly onto the ambulance floor
(Figure 69). Verify that the loading wheels are resting
on the ambulance floor, the safety bar is caught by the
safety hook,
and
that the loading-end legs have not
lifted off the ground.
2. Press the Mode-Select button four times to enter the
load-height set mode. See “Mode-Select Button” on
○
The background screen is white to indicate the load-
height set mode (Figure 70). The wheels are red to
indicate that the iN∫X must not be rolled.
○
A blue line is displayed below the loading wheel, and
a 15 second countdown timer is shown in the lower-
left corner of the display.
○
The symbol is displayed to indicate that pressing
the button will set a new loading height.
3. Press and hold to set the loading height before the
timer expires. When pressed, a new, 5-second timer
appears on the display. Continue holding until the
timer expires.
4. Turn the power switch OFF and then ON. See “Power
5. Retract and extend the legs to verify the iN∫X stops at
the new loading height. Test loading and unloading
the iN∫X with the ambulance.
Error Setting Loading Height
No loading height will be set when the iN∫X
is at the factory-set
maximum transport
height
or
maximum height
.
5
69
70
Initial Setup
Countdown Timer