![Ferno 24 MAXX User Manual Download Page 27](http://html1.mh-extra.com/html/ferno/24-maxx/24-maxx_user-manual_554363027.webp)
27
© Ferno-Washington, Inc 234-3307-04 May 2011
24•MAXX®, 24•miniMAXX®, 24H•miniMAXX®
Maintenance
7.7 lubricating the Cot
Disinfect and clean the cot before applying lubricant. Use
the lubricants designated below. Do not lubricate points
marked with the “Do Not Lubricate” symbol.
The lubrication points are the same for the Model 24-MAXX,
Model 24-miniMAXX, and Model 24H-miniMAXX cots.
Do not lubricate
lubricate
Leg Joints
4
Illustration at Left Shows Top-Down
View with Bed Surface Removed
Steel Rod
(inside spring)
3
I-Beam
2
1
1
1
Release-Handle
Hinges (7)
1
important
Lubricating parts that should not be lubricated allows
dirt and foreign particles to collect on those parts,
resulting in damage. Lubricate
only
the numbered
reference points shown.
lUbriCaTiOn pOinTs
(Use a small amount of lubricant)
Reference #
Location
# of Points
Lubricate With
Release Handle Hinges
7
WRL-191-S or EZ-1 synthetic lubricant
I-Beam
1
White lithium grease
Steel Rod (inside leg-return spring)
1
WRL-191-S or EZ-1 synthetic lubricant
Leg Joints
4
WRL-191-S or EZ-1 synthetic lubricant
Wheels (sealed bearings)
0
Do not lubricate
Upper Leg Pivot Points
0
Do not lubricate
1
3
2
4