Password (Pass) -
Select a number to act as password. This
password will be requested when entering the Configuration
Menu. To deactivate the password select “Off”.
Factory Settings (Fact) -
Factory default configuration. Se
-
lect “yES” to activate the factory default configuration.
Version (Ver) -
Firmware version installed.
Light (Ligh) -
Luminosity. Select between 5 predefined levels
of luminosity.
Tools Menu
Menu options OPT1 and OPT2 give access to the configura
-
tion menus of the installed options. The menu depends on the
installed option. If there is no option installed the instrument
shows “nonE”.
Options Menu
When the instrument detects that the displayed value does
not correspond to the input value, the display will flash and
alternate with a message.
“
h.udr
” Hardware underrange. The thermocouple signal is
lower than minimum readable signal (-30mV).
“
h.oVr
” Hardware overrange. The thermocouple signal is
higher than maximum readable signal (80mV).
“
d.udr
” Display underrange. The Pt100/RTD is shortcir-
cuited.
“
hoLd
” The instrument is showing the value present when
the hold function was activated. Hold function is active.
“
Min
”
The instrument displays the minimum displayed
value in memory. The minimum visualization is active.
“
MAX
” The instrument displays the maximum displayed
value in memory. The maximum visualization is active.
“
brk
”
The instrument displays “brk” with either “9999” or
“-9999” when : 1) resistance measured is higher than 390
Ohms (higher than 850ºC for PT100/RTD measure), or 2) in
case the third wire ohm is higher than 15 Ohm, or 3) thermo-
couple measure is open circuit.
Err.1
Password incorrect.
Err.2
The instrument has detected an installed option but
was unable to communicate.