User’s Manual S40-T

User’s Manual S40-T

FEMA ELECTRÓNICA - Page 3

Button

UP

Logo

Alarms

Units

Power

Signal

Option1

Option2

Button

SQ

Button

LE

Front View

Rear View

Earth connection - Although a terminal is offered for earth connection, the 

connection is optional. The instrument does not need this connection for 

correct functioning nor for compliance with the security regulations.
Fuse - To comply with security regulation 61010-1, add to the power line 

a protection fuse acting as disconnection element, easily accessible to the 

operator and identified as a protection device.

 

Power “H”   fuse 250mA time-lag

 

Power “L”   fuse 400mA time-lag

6 7 8 9 0

~

~

Power Connections

Input signal connections - Pt100/RTD

1 2 3 4 5

RTD/Pt100

2 wires

1 2 3 4 5

RTD/Pt100

3 wires

Measure can be selected for 2 or 3 wire systems.

Select input Pt100 2Wires in the 

menu. Terminal 2 can be left un-

connected.

Select input Pt100 3Wires in the 

menu

Input signal connections - Thermocouple

1 2 3 4 5

+

Thermocouple

-

To configure, select the appropriate thermocouple type in the 

configuration menu.

+

-

Summary of Contents for S40-T

Page 1: ...th internal CJC selectable enabled or disabled Selectable behavior in case of sensor break to_high or to_low Reduced housing 72x36mm DIN standard Display in 4 digits and negative sign Maximum and mini...

Page 2: ...on the specified protection level Disconnect the instrument from power before starting any maintenance and or installation action Risk of electrical shock Instrument terminals can be connected to dan...

Page 3: ...he power line a protection fuse acting as disconnection element easily accessible to the operator and identified as a protection device Power H fuse 250mA time lag Power L fuse 400mA time lag 6 7 8 9...

Page 4: ...able 3 Offset drift see table 3 Span drift see table 3 Note span drift includes also the offset drift CJC automatic On Off selectable CJC accuracy 1 0 C CJC thermal drift 0 04 C On sensor break to_hig...

Page 5: ...nu or leaves the Informa tion Menu Operating the Information Menu 36 8 32 5 Mechanical Dimensions mm 72 18 80 69 Panel cut out Configuration Conf Information on the configured input sensor and the deg...

Page 6: ...the menu Allows selec tion of the current option displayed During a value selection menu for example a setpoint value validates the value on display Button UP Moves vertically on the menu Displays the...

Page 7: ...tion See menu for Alarm1 See menu for Alarm1 See menu for Alarm1 Alarm1 Setpoint Hysteresis Delay Active On Inactive Off Max Maximum Min Minimum Alarm management Alarm as max Alarm as min Counts Secon...

Page 8: ...fected on the same way as the setpoint with hysteresis and delays Display Menu Functions on this menu allow for configuration of the visu alization Steps StEP Display changes on predefined steps Value...

Page 9: ...message h udr Hardware underrange The thermocouple signal is lower than minimum readable signal 30mV h oVr Hardware overrange The thermocouple signal is higher than maximum readable signal 80mV d udr...

Page 10: ...ber 2011 Daniel Junc Quality Manager All instruments are warranted against all manufacturing defects for a period of 24 MONTHS from the shipment date This warranty does not apply in case of misuse acc...

Page 11: ...User s Manual S40 T FEMA ELECTR NICA Page 11 T H I S P A G E B L A N K...

Page 12: ...ucts ELECTRONIC INSTRUMENTATION FOR INDUSTRY www fema es FEMA ELECTR NICA S A Altimira 14 Pol Ind Santiga E08210 Barber del Vall s BARCELONA SPAIN Tel 34 93 729 6004 www fema es Fax 34 93 729 6003 inf...

Reviews: