User’s Manual S40-T
User’s Manual S40-T
FEMA ELECTRÓNICA - Page 3
Button
UP
Logo
Alarms
Units
Power
Signal
Option1
Option2
Button
SQ
Button
LE
Front View
Rear View
Earth connection - Although a terminal is offered for earth connection, the
connection is optional. The instrument does not need this connection for
correct functioning nor for compliance with the security regulations.
Fuse - To comply with security regulation 61010-1, add to the power line
a protection fuse acting as disconnection element, easily accessible to the
operator and identified as a protection device.
Power “H” fuse 250mA time-lag
Power “L” fuse 400mA time-lag
6 7 8 9 0
~
~
Power Connections
Input signal connections - Pt100/RTD
1 2 3 4 5
RTD/Pt100
2 wires
1 2 3 4 5
RTD/Pt100
3 wires
Measure can be selected for 2 or 3 wire systems.
Select input Pt100 2Wires in the
menu. Terminal 2 can be left un-
connected.
Select input Pt100 3Wires in the
menu
Input signal connections - Thermocouple
1 2 3 4 5
+
Thermocouple
-
To configure, select the appropriate thermocouple type in the
configuration menu.
+
-