Fema LP4 User Manual Download Page 5

 Model LP4

SERIES SPECIAL . LOOP POWERED METERS

www.fema.es 

5

HOW TO ENTER THE ‘FAST ACCESS’ MENU

With the instrument in ‘

normal mode

’ of operation (see section 10), press 

the ‘

UP

’ (

5

) key. The instrument enters into the 

‘fast access

’ menu.

HOW TO OPERATE INSIDE THE ‘FAST ACCESS’ MENU

The available functions inside the ‘

fast access

’ menu can be configured 

(see  section  13.4).  By  default,  ‘

memory of maximum

’,  ‘

memory of 

minimum

’ and ‘

measure function

’ are available. See section 13.4 for a list 

of configurable functions.

 press the ‘

UP

’ (

5

) key to move to the next function.

 press the ‘

SQ

’ (

<

) key to activate the selected function.

When the function is active, the display remains flashing. Press any key 

to deactivate the function (display stops flashing). 

‘ROLLBACK’ FUNCTION

If there is no interaction from the operator for 30 seconds, the 

instrument exits the ‘

fast access

’ menu and returns to ‘

normal 

mode

’ of operation.

HOW TO EXIT ‘FAST ACCESS’ MENU

To exit the ‘

fast access

’ menu, press the ‘

LE

’ (

3

) key, or press the key 

UP

’ (

5

) key until the parameter ‘- - - -’ appears, and select by pressing 

the ‘

SQ

’ (

<

) key, or wait without pressing any key until the automatic 

rollback

’ activates.

When exiting the ‘

fast access

’ menu, the instrument returns to ‘

normal 

mode

’ of operation.

12. How to operate the ‘Fast access’ menu

Table 6 

Example of ‘Fast access’ menu with all functions set to ‘on’

‘Memory of max.’

‘Memory of min.’

Exit

Measured value

See section 13.4 for a list and a description of available functions.

HOW TO EXIT THE ‘CONFIGURATION MENU’

When  exiting  the  ‘

configuration  menu

’  without  changes  (either  by 

rollback

’ activation or because there are no changes in the configuration), 

the horizontal leds light down from top to bottom, and the instrument 

returns to ‘

normal mode

’ of operation.

When exiting the ‘

configuration menu

’ with changes, the display leds light 

a round shape while the new configuration is stored. When the round 

shape is finished, the horizontal leds light down from top to bottom and 

the instrument restarts with the new configuration. The instrument is 

now in ‘

normal mode

’ of operation.

‘ROLLBACK’ FUNCTION

If there is no interaction from the operator for 30 seconds, the instrument 

exits the ‘

configuration menu

’ discarding changes, and returns to ‘

normal 

mode

’ of operation.

Summary of Contents for LP4

Page 1: ...www fema es SPECIAL YOUR METER APPLICATIONS Model LP4 USER S MANUAL Loop mA LOOP POWERED PROCESS METER SERIES SPECIAL 5137r01...

Page 2: ...6 13 4Key UP fast access 7 13 5 Tools menu 7 14 Technical specifications 8 15 Error codes 8 16 Mechanical dimensions mm inch 8 17 Precautions on installation 9 18 Warranty 9 19 Factory default paramet...

Page 3: ...d connections Inverted connections will not damage the instrument the loop will remain open and the instrument does not operate display does not light Terminal 2 is not connected Connecting any of the...

Page 4: ...ries parameters and select configuration values Key SQ functions as the ENTER key It selects the menu entry currently displayed At numerical value entries it validates the number displayed Key UP 5 mo...

Page 5: ...r appears and select by pressing the SQ key or wait without pressing any key until the automatic rollback activates When exiting the fast access menu the instrument returns to normal mode of operation...

Page 6: ...ads wait while performs the reading of the actual input signal value and associates this value to the input low parameter When finished the instrument goes back to the Correction low F Lo menu entry a...

Page 7: ...ect Measure function MEAS to activate an access to visualize the actual input signal value in milliamperes without the scaling 13 5 Tools menu The Tools tool menu groups several functions at the Input...

Page 8: ...isplay display refresh 2 times second acquisition per second 2 acquisitions second memory retention values for maximum and minimum reading are stored every 10 minutes INPUT SIGNAL signals accepted 4 2...

Page 9: ...r than the minimum current needed The instrument is designed to be panel mounted An appropriate ventilation of the instrument must be assured Do not expose the instrument to excess of humidity Maintai...

Page 10: ...of LP4 loop powered meter placed close to the transmitter using the existing loop to the PLC 4 20mA transmitter Units LP4 loop powered meter PLC The LP4 loop powered meter can be directly inserted int...

Page 11: ...2 3Protector KIP Wall mount housing Together with the KIP protector offers a full IP65 protection For for 96x48mm panel meters 22 2 4WME wall housing Benchtop housing for for 96x48mm panel meters Han...

Page 12: ......

Reviews: