Fema C Series User Manual Download Page 27

FEMA ELECTRÓNICA . Series C . C40-D

27

Protocol  

 

Modbus RTU

Bus type  

 

RS-485, configurable for 9.600 or 4800 bps

Addresses 

 

1 to 247

Formats  

 

configurable 8n1, 8e1, 8o1, 8n2

Function code   

read register

Registers 

 

reading value (16 bits)

 

 

 

 

number of decimals (16 bits)

Errors   

 

function 

not 

supported

 

 

 

 

register not accessible

Isolation  

 

1000 Vdc

Slots allowed 

 

Opt.1

2.3 Module S1 (Modbus RTU)

The S1 module offers 1 Modbus RTU serial output, isolated, to be 

installed at slot Opt1.
Configuration is done from the front keypad, through the configura

-

tion menu. The instrument must be informed that there is a Mod

-

bus RTU module at slot.1 and this is done at the configuration menu 

‘Tool’ \ ‘out.1’ 

(see section 1.29.7). 

Then configure the bus param

-

eters at the configuration menu ‘out.1’ \ ‘r485’ 

(see section 1.29.8).

The S1 module can be ordered installed in to a Series C instrument 

or standalone for later installation, as it does not require soldering or 

special configuration.

Terminal B 

B  

Terminal A 

A  

Terminal G 

GND

Power

Signal

Installing, calibrating and configuring a M1 module

If the M1 module has been acquired installed in the instrument, then 

the module has been factory calibrated. In this case you can jump 

directly to point 7 below.
If the M1 module has been acquired separately and installation is 

needed, follow the next steps :

1.

 Access the instrument and install the M1 module at slot Opt.1 

(see section 1.26).

2.

 From the ‘Configuration menu’ go to ‘tool’ \ ‘out.1’ and select 

value ‘420’ 

(see section 1.29.7) 

to inform the instrument that there 

is a M1 analog output module installed.

From this point you will need the calibration sheet provided together 

with  the  module  M1.  This  sheet  shows  the  values  for  parameters 

‘cAL.y / 4mA’ and ‘cAL.y / 20mA’.

3.

 When selecting ‘420’ the instrument shows ‘

cAL.n

’. (‘

Do not cali-

brate

’). Press ‘UP’ (

5

) to read ‘

cAL.y

’ (‘

Calibrate

’).

4.

 When selecting ‘

cAL.y

’ (‘

Calibrate

’) select ‘4 mA’. Display shows 

a number. Introduce the value of the ‘cAL.y / 4mA’ indicated at the  

calibration sheet provided together with the M1 module. Increase 

the numerical value with key ‘UP’ (

5

) and reduce the value with 

key ‘LE’ (

3

). Validate the value with key ‘SQ’ (

<

). See below an 

image of the configuration menu.

5. 

Repeat with the ‘20 mA’ parameter, and update with the value 

‘cAL.y / 20mA’ indicated at the  calibration sheet. Validate with key 

‘SQ’ (

<

), and press key ‘LE’ (

3

) several times to exit the ‘Configu

-

ration menu’.

6. 

The instrument will reboot when exiting the menu. The analog 

output is calibrated.
Note : if you do not have the calibration sheet which was attached 

to the M1 module at manufacturing time, you can use a miliamme

-

ter to measure the mA at the output loop. Then access parameters 

‘cAL.y / 4mA’ and ‘cAL.y / 20mA’ and adjust the calibration value by 

checking the miliammeter value.

At this point the module is installed and calibrated. Now configure 

the reading associated to the output signal 4 mA and 20 mA.

7. 

Configure the reading associated to the 4 mA output signal and 

the reading associated to the 20 mA output signal, at the configu

-

ration menu ‘out.1’ \ ‘420’ 

(see section 1.29.8).

Do not calibrate

Calibrate

Option 1

analog output

relay

Disabled

Tools

Modbus RTU

B A G

Opt.1

GND

B

A

Summary of Contents for C Series

Page 1: ...ith 4 digits 9999 1999 Fast access function to alarm setpoints external control to activate predefined functions Eco mode for low consumption selectable dual scaling 5 levels of configurable brightnes...

Page 2: ...23 Eco mode 13 1 24 External control 13 1 25 Second scaling 13 1 26 To open the instrument 14 1 27 How to operate the menus 15 1 28 Messages and errors 15 1 29 Configuration menu 16 1 29 1 Input signa...

Page 3: ...terminals 1 and 4 Signals for 5 A current AC and DC must be con nected at terminals 3 and 4 All other signals must be connected between terminals 2 and 4 Terminal 5 is a multifunction terminal config...

Page 4: ...r process signals 5 Vdc for potentiometers at terminal 5 see sections 1 5 and 1 16 Power supply power U 18 to 265 Vac dc isolation 1500 Veff isolation tested for 60 sec AC frequency 50 60 Hz consumpti...

Page 5: ...Jumpers S Jumper T AC voltages and currents 600 Vac G I 4 5 200 Vac I 20 Vac A I 2 Vac B I 200 mVac C I 60 mVac E I 5 Aac I 20 mAac D I DC voltages and currents 600 Vdc G 4 5 200 Vdc 20 Vdc A 2 Vdc B...

Page 6: ...steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vac ranges Veff Scale by default Scalable Jumper S see section 1...

Page 7: ...ion 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vdc ranges Scale by default Scalable Jumper S see section 1 9 Jumper T see section 1...

Page 8: ...ouple K E 4 5 100 1350 C 148 2462 F 2 tc 4 tc 3 Thermocouple J 100 1200 C 148 2192 F Thermocouple E 100 1000 C 148 1832 F Thermocouple N 100 1300 C 148 2372 F Thermocouple L 100 900 C 148 1652 F Therm...

Page 9: ...0 see section 1 29 7 This pa rameter is associated to the specific model of probe installed Sensor Jumper S see section 1 9 Jumper T see section 1 9 Range in C in F Total error Connection terminals Cu...

Page 10: ...n for internal jump ers is indicated at section 1 9 See below connections for NTC probe 1 14 Measuring with NTC probes NTC R25 configurable Jumper S see section 1 9 Jumper T see section 1 9 Range of m...

Page 11: ...nnection terminals Zin 4 20 mA passive needs Vexc 0 100 0 de 9999 a 1999 D 1 2 0 15 25 mA 2 signal 5 Vexc 4 7Ohm active 4 5 2 mA 4 mA 0 10 Vdc passive needs Vexc A 1 2 0 20 25 Vdc 2 Vdc 4 comm 5 Vexc...

Page 12: ...n 1 9 for a list of available functions Start up connections and jumpers For instrument start up follow the steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for int...

Page 13: ...ree potential contact called Exter nal control To configure the second scaling 1 configure the internal jumper T to activate the EK external control function see section 1 9 2 associate the external c...

Page 14: ...X and then insert clips A B C and D in this order The internal structure of the instrument is shown in the graphic be low Module Opt 2 connects to module Opt 1 Module Opt 1 connects to the display Opt...

Page 15: ...t validates the num ber displayed Key UP 5 the UP 5 key gives access to the fast access menu Inside the menus it moves vertically through the different menu op tions At menus with numerical value entr...

Page 16: ...able options Pt100 3 wire Pt100 2 wire Pt500 Pt1000 Ni100 Ni200 and Ni1000 For measures with 2 wires a manual compensation for the wire resistance error is available us ing parameter Offset reading oF...

Page 17: ...ximum type alarm or minimum type activates when the dis play value is higher or lower than the setpoint value at the Setpoint SEt parameter enter the value for the alarm activation point This paramete...

Page 18: ...vate a tare to the signal The actual signal will read 0 The calibration is not changed select parameter Decimal point 0 dP 0 to remove the decimal point on display select parameter Decimal point 1 dP...

Page 19: ...compen sate for the possible error introduced by the resistance of the signal wires when measuring Pt probes Ni probes or resistances For more information see section 1 22 1 29 7 Menu Tools Option 1...

Page 20: ...Ni1000 PTC and NTC For thermocouple probes resolution is fixed to 1 at the Degrees dEG menu select the temperature values to be read in C degrees Celsius or in F degrees Fahrenheit at the Alpha ALPh...

Page 21: ...r 485 at Tools out 1 r 485 to have the serial output configuration at this menu entry Configure the address for the instrument parameter Address 1 to 247 Addr the bus speed parameter Speed kbps bAud...

Page 22: ...c range 2 Vdc range 200 mVdc range 60 mVdc range 5 Adc range 20 mAdc range Process 4 20 mA range 0 10 Vdc range Thermocouples thermocouple K thermocouple J thermocouple E thermocouple N thermocouple L...

Page 23: ...ss Memory of max Memory of min Setpoint 1 Setpoint 2 Degrees degrees Celsius degrees Fahrenheit Temperature resolution 1 resolution 0 1 resolution Second scaling Decimal point Display Low Display High...

Page 24: ...elected at Tool Out 1 Analog output Display Low Display High Modbus RTU Address 1 to 247 Speed Kbps Format Password Version Factory reset Minimum Brightness Maximum Standard Alpha Alpha 385 Alpha 390...

Page 25: ...ent Industrial CISPR 11 Instrument Class A Class B Group 1 For a detailed declaration see document www fema es docs 4152_CE Declaration_C40 D_en pdf Barber del Vall s January 2017 Xavier Junc Product...

Page 26: ...portional to the reading of the instrument The mA output can be connected to work in active loop the module provides the power of the loop or passive mode the power of the loop is not provided by the...

Page 27: ...420 see section 1 29 7 to inform the instrument that there is a M1 analog output module installed From this point you will need the calibration sheet provided together with the module M1 This sheet s...

Page 28: ...the front filter Inter nal press buttons for configuration are accessible Optionally request the instrument preconfigured from factory Without front keypad 3 1 Option NBT O ring for IP65 front protec...

Page 29: ...ll mount housing Together with the KIP protector offers a full IP65 protection For Series C of panel meters 4 4 WME wall housing Benchtop housing for Series C and of panel meters Handle with three sel...

Page 30: ...FEMA ELECTR NICA Series C C40 D 30 Notes...

Page 31: ...FEMA ELECTR NICA Series C C40 D 31 Notes...

Page 32: ......

Reviews: