Fema C Series User Manual Download Page 10

FEMA ELECTRÓNICA . Series C . C40-D

10

 Accepted NTC probes

The NTC probe is a temperature variable resistor with a 

temperature - resistance curve defined by two param

-

eters called ‘R25’ and ‘beta’. By default, the instrument 

is configured for a standard NTC with ‘R25’ of 10K and 

‘beta’  of  3500.  The  measured  temperature  ranges  from  -60 ºC  to 

150 ºC. 
Different NTC probes can be used, by configuring the instrument for 

different values of ‘R25’ and ‘beta’. The instrument measures resis

-

tance from 100 R up to 1 MOhm. Use 

‘Table 12’

 to write down values 

for your NTC.

 Parameters ‘R

25

’ and ‘Beta’

The instrument accepts values of ‘beta’ between 2500 and 5500, and 

values of R25 between 1.0 K and 200.0 K

 Resolution and units

The temperature resolution using NTC temperature probes is con

-

figurable to 1º or 0.1º. Reading can be configured in ºC (degrees Cel

-

sius) or ºF (degrees Fahrenheit).

Range  

of measure

NTC probe

Temp.

R25

Beta

100 R

1 MR

Table 12 - Data from NTC datasheet

 

Sensor break detection

In case of sensor break, the instrument will show ‘

h.ovr

’ or ‘

h.udr

’ 

(see section 1.28) 

depending on the broken cable.

 

Terminal 5 ‘multifunction’ - External control

Terminal 5 remains configured as ‘EK’ external control function. See 

section 

1.9 

for a list of available functions.

 

Start-up, connections and jumpers

For instrument start-up follow the steps listed at section 

1.7

Signal 

connections are indicated at section 

1.5.

 Location for internal jump

-

ers is indicated at section 

1.9.

See below connections for NTC probe :

1.14 Measuring with NTC probes

NTC ‘R

25

’* 

(

configurable)

Jumper ‘S’ 

(see section 1.9)

Jumper ‘T’ 

(see section 1.9)

Range  

of measure

Accuracy 

(% of reading)

Connection

 

(

terminals)

Beta* 

(configurable)

10K

F K

4-5

from -60 ºC  

to 150 ºC

<1.5% of 

reading

2 (NTC +) 

4 (NTC -)

3500

Table 13 - *‘Beta’ is configurable from 2000 to 5500. ‘R25’ is configurable from 1.0 K up to 200.0 K.

ntc +

ntc -

1 2 3 4 5

1.15 Measuring with PTC probes

 Accepted PTC probes

The  instrument  accepts  direct  connection  of  PTC 

probes. Accepted PTC probes are listed a

‘Table 14’.

 Resolution and units

The  temperature  resolution  using  PTC  temperature  probes  is  con

-

figurable to 1º or 0.1º. Reading can be configured in ºC (degrees Cel

-

sius) or ºF (degrees Fahrenheit).

Family

Sensor

Jumper ‘S’ 

(see section 1.9)

Jumper ‘T’ 

(see section 1.9)

Range in ºC

 

(in ºF)

Total error

KTY 121 KTY81-121 

KTY82-121

F

4-5

-55 / 150 ºC 

(-67 / 302 ºF)

<1º

KTY 210 KTY81-210 

KTY82-210

FHK

KTY 220 KTY81-220 

KTY82-220

FHK

Table 14 - Ranges of temperature for PTC probes

 

Sensor break detection

In case of sensor break, the instrument will show ‘

h.ovr

’ or ‘

h.udr

’ 

(see section 1.28) 

depending on the broken cable.

 

Terminal 5 ‘multifunction’ - External control

Terminal 5 remains configured as ‘EK’ external control function. See 

section 

1.9 

for a list of available functions.

ptc +

ptc -

1 2 3 4 5

 

Start-up, connections and jumpers

For instrument start-up follow the steps listed at section 

1.7

Signal connections are indicated at section 

1.5.

 Loca

-

tion for internal jumpers is indicated at section 

1.9.

See below connections for PTC probe :

Summary of Contents for C Series

Page 1: ...ith 4 digits 9999 1999 Fast access function to alarm setpoints external control to activate predefined functions Eco mode for low consumption selectable dual scaling 5 levels of configurable brightnes...

Page 2: ...23 Eco mode 13 1 24 External control 13 1 25 Second scaling 13 1 26 To open the instrument 14 1 27 How to operate the menus 15 1 28 Messages and errors 15 1 29 Configuration menu 16 1 29 1 Input signa...

Page 3: ...terminals 1 and 4 Signals for 5 A current AC and DC must be con nected at terminals 3 and 4 All other signals must be connected between terminals 2 and 4 Terminal 5 is a multifunction terminal config...

Page 4: ...r process signals 5 Vdc for potentiometers at terminal 5 see sections 1 5 and 1 16 Power supply power U 18 to 265 Vac dc isolation 1500 Veff isolation tested for 60 sec AC frequency 50 60 Hz consumpti...

Page 5: ...Jumpers S Jumper T AC voltages and currents 600 Vac G I 4 5 200 Vac I 20 Vac A I 2 Vac B I 200 mVac C I 60 mVac E I 5 Aac I 20 mAac D I DC voltages and currents 600 Vdc G 4 5 200 Vdc 20 Vdc A 2 Vdc B...

Page 6: ...steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vac ranges Veff Scale by default Scalable Jumper S see section 1...

Page 7: ...ion 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vdc ranges Scale by default Scalable Jumper S see section 1 9 Jumper T see section 1...

Page 8: ...ouple K E 4 5 100 1350 C 148 2462 F 2 tc 4 tc 3 Thermocouple J 100 1200 C 148 2192 F Thermocouple E 100 1000 C 148 1832 F Thermocouple N 100 1300 C 148 2372 F Thermocouple L 100 900 C 148 1652 F Therm...

Page 9: ...0 see section 1 29 7 This pa rameter is associated to the specific model of probe installed Sensor Jumper S see section 1 9 Jumper T see section 1 9 Range in C in F Total error Connection terminals Cu...

Page 10: ...n for internal jump ers is indicated at section 1 9 See below connections for NTC probe 1 14 Measuring with NTC probes NTC R25 configurable Jumper S see section 1 9 Jumper T see section 1 9 Range of m...

Page 11: ...nnection terminals Zin 4 20 mA passive needs Vexc 0 100 0 de 9999 a 1999 D 1 2 0 15 25 mA 2 signal 5 Vexc 4 7Ohm active 4 5 2 mA 4 mA 0 10 Vdc passive needs Vexc A 1 2 0 20 25 Vdc 2 Vdc 4 comm 5 Vexc...

Page 12: ...n 1 9 for a list of available functions Start up connections and jumpers For instrument start up follow the steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for int...

Page 13: ...ree potential contact called Exter nal control To configure the second scaling 1 configure the internal jumper T to activate the EK external control function see section 1 9 2 associate the external c...

Page 14: ...X and then insert clips A B C and D in this order The internal structure of the instrument is shown in the graphic be low Module Opt 2 connects to module Opt 1 Module Opt 1 connects to the display Opt...

Page 15: ...t validates the num ber displayed Key UP 5 the UP 5 key gives access to the fast access menu Inside the menus it moves vertically through the different menu op tions At menus with numerical value entr...

Page 16: ...able options Pt100 3 wire Pt100 2 wire Pt500 Pt1000 Ni100 Ni200 and Ni1000 For measures with 2 wires a manual compensation for the wire resistance error is available us ing parameter Offset reading oF...

Page 17: ...ximum type alarm or minimum type activates when the dis play value is higher or lower than the setpoint value at the Setpoint SEt parameter enter the value for the alarm activation point This paramete...

Page 18: ...vate a tare to the signal The actual signal will read 0 The calibration is not changed select parameter Decimal point 0 dP 0 to remove the decimal point on display select parameter Decimal point 1 dP...

Page 19: ...compen sate for the possible error introduced by the resistance of the signal wires when measuring Pt probes Ni probes or resistances For more information see section 1 22 1 29 7 Menu Tools Option 1...

Page 20: ...Ni1000 PTC and NTC For thermocouple probes resolution is fixed to 1 at the Degrees dEG menu select the temperature values to be read in C degrees Celsius or in F degrees Fahrenheit at the Alpha ALPh...

Page 21: ...r 485 at Tools out 1 r 485 to have the serial output configuration at this menu entry Configure the address for the instrument parameter Address 1 to 247 Addr the bus speed parameter Speed kbps bAud...

Page 22: ...c range 2 Vdc range 200 mVdc range 60 mVdc range 5 Adc range 20 mAdc range Process 4 20 mA range 0 10 Vdc range Thermocouples thermocouple K thermocouple J thermocouple E thermocouple N thermocouple L...

Page 23: ...ss Memory of max Memory of min Setpoint 1 Setpoint 2 Degrees degrees Celsius degrees Fahrenheit Temperature resolution 1 resolution 0 1 resolution Second scaling Decimal point Display Low Display High...

Page 24: ...elected at Tool Out 1 Analog output Display Low Display High Modbus RTU Address 1 to 247 Speed Kbps Format Password Version Factory reset Minimum Brightness Maximum Standard Alpha Alpha 385 Alpha 390...

Page 25: ...ent Industrial CISPR 11 Instrument Class A Class B Group 1 For a detailed declaration see document www fema es docs 4152_CE Declaration_C40 D_en pdf Barber del Vall s January 2017 Xavier Junc Product...

Page 26: ...portional to the reading of the instrument The mA output can be connected to work in active loop the module provides the power of the loop or passive mode the power of the loop is not provided by the...

Page 27: ...420 see section 1 29 7 to inform the instrument that there is a M1 analog output module installed From this point you will need the calibration sheet provided together with the module M1 This sheet s...

Page 28: ...the front filter Inter nal press buttons for configuration are accessible Optionally request the instrument preconfigured from factory Without front keypad 3 1 Option NBT O ring for IP65 front protec...

Page 29: ...ll mount housing Together with the KIP protector offers a full IP65 protection For Series C of panel meters 4 4 WME wall housing Benchtop housing for Series C and of panel meters Handle with three sel...

Page 30: ...FEMA ELECTR NICA Series C C40 D 30 Notes...

Page 31: ...FEMA ELECTR NICA Series C C40 D 31 Notes...

Page 32: ......

Reviews: