Fema C Series User Manual Download Page 12

FEMA ELECTRÓNICA . Series C . C40-D

12

Ranges 

 

of measure

Scale 

 

by default

Scalable

Jumper ‘S’ 

(see section 1.9)

Jumper ‘T’ 

(see section 1.9)

Accuracy 

(% of reading)

0 to 5 K

5.000

from 9999  

to -1999

F H K

4-5

<1.5% of 

reading

0 to 50 K

50.00

F K

Table 17 - Ranges of measure for resistances

• Measuring resistive signals

The instrument accepts the measure of resistances and 

provides two ranges of measure from 0 to 5 K and from  

0 to 50 K.

 

Compensating the resistance of the signal wire

Resistances are measured with 2 wires system. To compensate for 

the possible error introduced by the resistance of the signal wires, 

the instrument allows to configure a fixed number of counts to be 

added to the reading, both in positive or negative. This is done with 

the parameter ‘

Offset reading

’ (‘

oFFS

’) 

(see section 1.22).

 

Scaling

The instrument allows to scale the reading to 4 digits (9999 / -1999) 

with configurable  decimal point to any position. The ‘

second scaling

’ 

function can also be used 

(see section 1.25)

1.18 Measuring resistances

 

Response times

The response time to a signal step is 300 mSeconds, independent of 

the signal range selected.

 

Terminal 5 ‘multifunction’ - External control

Terminal 5 remains configured as ‘EK’ external control function. See 

section 

1.9 

for a list of available functions.

 

Start-up, connections and jumpers

For instrument start-up follow the steps listed at section 

1.7

Signal 

connections are indicated at section 

1.5.

 Location for internal jump

-

ers is indicated at section 

1.9.

See below connections for resistance measures :

res +

res -

1 2 3 4 5

Nominal

 

pot. value

Ranges 

 

of measure

Scale 

 

by default

Scalable

Jumper ‘S’ 

(see section 1.9)

Jumper ‘T’ 

(see section 1.9)

Accuracy 

(% FS)

from 500 R  

up to 20 K

0 to 100 %

0/100.0

from 9999  

to -1999

A

2-3

<0.5%

Table 18 - Ranges of measure for potentiometers

• Measuring potentiometers

The instrument accepts the measure of 3 wire potenti

-

ometers, with a single default range of 0/100 % of the 

potentiometer span.

 

Scaling

The instrument allows to scale the reading to 4 digits (9999 / -1999) 

with configurable . 

 

Response times

The response time to a signal step is 300 mSeconds, independent of 

the signal range selected.

 

Terminal 5 ‘multifunction’ - ‘Vexc’

To measure potentiometer signals, set internal jumper ‘T’ at position 

2-3 

(see section 1.9). 

Terminal 5 will be assigned to a +5 Vdc excita

-

tion voltage for the potentiometer.

1.19 Measuring potentiometers

 

Start-up, connections and jumpers

For instrument start-up follow the steps listed at section 

1.7

Signal 

connections are indicated at section 

1.5.

 Location for internal jump

-

ers is indicated at section 

1.9.

See below connections for measure of potentiometers ::

Signal

Pot- (comm.)

Pot+ (5 Vdc)

1 2 3 4 5

Summary of Contents for C Series

Page 1: ...ith 4 digits 9999 1999 Fast access function to alarm setpoints external control to activate predefined functions Eco mode for low consumption selectable dual scaling 5 levels of configurable brightnes...

Page 2: ...23 Eco mode 13 1 24 External control 13 1 25 Second scaling 13 1 26 To open the instrument 14 1 27 How to operate the menus 15 1 28 Messages and errors 15 1 29 Configuration menu 16 1 29 1 Input signa...

Page 3: ...terminals 1 and 4 Signals for 5 A current AC and DC must be con nected at terminals 3 and 4 All other signals must be connected between terminals 2 and 4 Terminal 5 is a multifunction terminal config...

Page 4: ...r process signals 5 Vdc for potentiometers at terminal 5 see sections 1 5 and 1 16 Power supply power U 18 to 265 Vac dc isolation 1500 Veff isolation tested for 60 sec AC frequency 50 60 Hz consumpti...

Page 5: ...Jumpers S Jumper T AC voltages and currents 600 Vac G I 4 5 200 Vac I 20 Vac A I 2 Vac B I 200 mVac C I 60 mVac E I 5 Aac I 20 mAac D I DC voltages and currents 600 Vdc G 4 5 200 Vdc 20 Vdc A 2 Vdc B...

Page 6: ...steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vac ranges Veff Scale by default Scalable Jumper S see section 1...

Page 7: ...ion 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vdc ranges Scale by default Scalable Jumper S see section 1 9 Jumper T see section 1...

Page 8: ...ouple K E 4 5 100 1350 C 148 2462 F 2 tc 4 tc 3 Thermocouple J 100 1200 C 148 2192 F Thermocouple E 100 1000 C 148 1832 F Thermocouple N 100 1300 C 148 2372 F Thermocouple L 100 900 C 148 1652 F Therm...

Page 9: ...0 see section 1 29 7 This pa rameter is associated to the specific model of probe installed Sensor Jumper S see section 1 9 Jumper T see section 1 9 Range in C in F Total error Connection terminals Cu...

Page 10: ...n for internal jump ers is indicated at section 1 9 See below connections for NTC probe 1 14 Measuring with NTC probes NTC R25 configurable Jumper S see section 1 9 Jumper T see section 1 9 Range of m...

Page 11: ...nnection terminals Zin 4 20 mA passive needs Vexc 0 100 0 de 9999 a 1999 D 1 2 0 15 25 mA 2 signal 5 Vexc 4 7Ohm active 4 5 2 mA 4 mA 0 10 Vdc passive needs Vexc A 1 2 0 20 25 Vdc 2 Vdc 4 comm 5 Vexc...

Page 12: ...n 1 9 for a list of available functions Start up connections and jumpers For instrument start up follow the steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for int...

Page 13: ...ree potential contact called Exter nal control To configure the second scaling 1 configure the internal jumper T to activate the EK external control function see section 1 9 2 associate the external c...

Page 14: ...X and then insert clips A B C and D in this order The internal structure of the instrument is shown in the graphic be low Module Opt 2 connects to module Opt 1 Module Opt 1 connects to the display Opt...

Page 15: ...t validates the num ber displayed Key UP 5 the UP 5 key gives access to the fast access menu Inside the menus it moves vertically through the different menu op tions At menus with numerical value entr...

Page 16: ...able options Pt100 3 wire Pt100 2 wire Pt500 Pt1000 Ni100 Ni200 and Ni1000 For measures with 2 wires a manual compensation for the wire resistance error is available us ing parameter Offset reading oF...

Page 17: ...ximum type alarm or minimum type activates when the dis play value is higher or lower than the setpoint value at the Setpoint SEt parameter enter the value for the alarm activation point This paramete...

Page 18: ...vate a tare to the signal The actual signal will read 0 The calibration is not changed select parameter Decimal point 0 dP 0 to remove the decimal point on display select parameter Decimal point 1 dP...

Page 19: ...compen sate for the possible error introduced by the resistance of the signal wires when measuring Pt probes Ni probes or resistances For more information see section 1 22 1 29 7 Menu Tools Option 1...

Page 20: ...Ni1000 PTC and NTC For thermocouple probes resolution is fixed to 1 at the Degrees dEG menu select the temperature values to be read in C degrees Celsius or in F degrees Fahrenheit at the Alpha ALPh...

Page 21: ...r 485 at Tools out 1 r 485 to have the serial output configuration at this menu entry Configure the address for the instrument parameter Address 1 to 247 Addr the bus speed parameter Speed kbps bAud...

Page 22: ...c range 2 Vdc range 200 mVdc range 60 mVdc range 5 Adc range 20 mAdc range Process 4 20 mA range 0 10 Vdc range Thermocouples thermocouple K thermocouple J thermocouple E thermocouple N thermocouple L...

Page 23: ...ss Memory of max Memory of min Setpoint 1 Setpoint 2 Degrees degrees Celsius degrees Fahrenheit Temperature resolution 1 resolution 0 1 resolution Second scaling Decimal point Display Low Display High...

Page 24: ...elected at Tool Out 1 Analog output Display Low Display High Modbus RTU Address 1 to 247 Speed Kbps Format Password Version Factory reset Minimum Brightness Maximum Standard Alpha Alpha 385 Alpha 390...

Page 25: ...ent Industrial CISPR 11 Instrument Class A Class B Group 1 For a detailed declaration see document www fema es docs 4152_CE Declaration_C40 D_en pdf Barber del Vall s January 2017 Xavier Junc Product...

Page 26: ...portional to the reading of the instrument The mA output can be connected to work in active loop the module provides the power of the loop or passive mode the power of the loop is not provided by the...

Page 27: ...420 see section 1 29 7 to inform the instrument that there is a M1 analog output module installed From this point you will need the calibration sheet provided together with the module M1 This sheet s...

Page 28: ...the front filter Inter nal press buttons for configuration are accessible Optionally request the instrument preconfigured from factory Without front keypad 3 1 Option NBT O ring for IP65 front protec...

Page 29: ...ll mount housing Together with the KIP protector offers a full IP65 protection For Series C of panel meters 4 4 WME wall housing Benchtop housing for Series C and of panel meters Handle with three sel...

Page 30: ...FEMA ELECTR NICA Series C C40 D 30 Notes...

Page 31: ...FEMA ELECTR NICA Series C C40 D 31 Notes...

Page 32: ......

Reviews: