Fema C Series User Manual Download Page 18

FEMA ELECTRÓNICA . Series C . C40-D

18

1.29 Configuration menu (cont.)

Memory of min.

Display hold

Memory of max.

External control

Decimal point 1

Decimal point 2

Decimal point 0

Second scaling

Decimal point 3

Disabled

An external on / off control can be connected to the rear of the in

-

strument. The operator can then control the activation of a config

-

ured function based on the state of this control. Function remains ac

-

tivated while the external contact is closed, and will deactivate when 

contact is open. To configure the function associated to the external 

control, enter the menu ‘

External control

’ (‘

EXt.c

’).

 select ‘

Disabled

’ (‘

oFF

’) to have no function associated.

 select parameter ‘

Memory of maximum

’ (‘

MAX

’) to visualize on 

display the memory of maximum reading.

 select parameter ‘

Memory of minimum

’ (‘

MIn

’) to visualize on 

display the memory of minimum reading.

 select parameter ‘

Display hold

’ (‘

hoLd

’) to maintain the actual 

reading ‘on hold’. Open the external contact to free the reading.

Parameters ‘

Memory of maximum

’ (‘

MAX

’), ‘

Memory of minimum

’ 

(‘

Min

’) and ‘

Display hold

’ (‘

hoLd

’) show on display a value which is 

not  the  actual  measured  input  signal,  therefor  the  values  for  this 

parameters will be shown in flash mode, indicating that the actual 

value is not the actual input value.

 select ‘

Tare function

’ (‘

tArE

’) to activate a tare to the signal. The 

actual signal will read ‘0’. The calibration is not changed.

 select parameter ‘

Decimal point 0

’ (‘

dP.0

’) to remove the decimal 

point on display. 

 select parameter ‘

Decimal point 1

’ (‘

dP.1

’) to activate the decimal 

point at position XXX.X

1.29.6 External control

The key ‘UP’ (

5

) at the front of the instrument gives access to a list 

of functions configurable by the operator. See section 

1.27

 for an ex

-

planation on how to operate the  ‘fast access’ menu. 
The ‘

Key UP (Fast access)

’ (‘

K.uP

’) menu allows to select which func

-

tions will be accessible when pressing the front key ‘UP’ (

5

). Select 

on

’ to activate each function.

 the ‘

Setpoint

1

’ (‘

ALr1

’) function allows to visualize and modify 

the setpoint value of alarm

1.

 the ‘

Setpoint

2

’ (‘

ALr2

’) function allows to visualize and modify 

the setpoint value of alarm

2.

the  ‘

Memory  of  maximum

’  (‘

MAX

’)

 

or

 

Memory  of  minimum

’ 

(‘

MIn

’) allows to visualize the memory of maximum and minimum 

reading. The memory resets when the instrument restarts (power 

off-on cycle, or when leaving the configuration menu with a change 

to be applied).

 the ‘

Tare function

’ (‘

tArE

’) allows to visualize the actual tare value.

1.29.4 Fast access

If only a single function is selected for the ‘fast access’ menu, press

-

ing the the ‘UP’ (

5

) key will shortly display the function name and 

then automatically jump to the function value.

1.29.5 Super fast access

Key UP 

(fast access)

Memory of max.

Memory of min.

Setpoint 1

Setpoint 2

Setpoint

Hystereses

Alarm type

Alarm 2

‘Tare’ function

‘Tare’ function

Summary of Contents for C Series

Page 1: ...ith 4 digits 9999 1999 Fast access function to alarm setpoints external control to activate predefined functions Eco mode for low consumption selectable dual scaling 5 levels of configurable brightnes...

Page 2: ...23 Eco mode 13 1 24 External control 13 1 25 Second scaling 13 1 26 To open the instrument 14 1 27 How to operate the menus 15 1 28 Messages and errors 15 1 29 Configuration menu 16 1 29 1 Input signa...

Page 3: ...terminals 1 and 4 Signals for 5 A current AC and DC must be con nected at terminals 3 and 4 All other signals must be connected between terminals 2 and 4 Terminal 5 is a multifunction terminal config...

Page 4: ...r process signals 5 Vdc for potentiometers at terminal 5 see sections 1 5 and 1 16 Power supply power U 18 to 265 Vac dc isolation 1500 Veff isolation tested for 60 sec AC frequency 50 60 Hz consumpti...

Page 5: ...Jumpers S Jumper T AC voltages and currents 600 Vac G I 4 5 200 Vac I 20 Vac A I 2 Vac B I 200 mVac C I 60 mVac E I 5 Aac I 20 mAac D I DC voltages and currents 600 Vdc G 4 5 200 Vdc 20 Vdc A 2 Vdc B...

Page 6: ...steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vac ranges Veff Scale by default Scalable Jumper S see section 1...

Page 7: ...ion 1 7 Signal connections are indicated at section 1 5 Location for internal jump ers is indicated at section 1 9 Vdc ranges Scale by default Scalable Jumper S see section 1 9 Jumper T see section 1...

Page 8: ...ouple K E 4 5 100 1350 C 148 2462 F 2 tc 4 tc 3 Thermocouple J 100 1200 C 148 2192 F Thermocouple E 100 1000 C 148 1832 F Thermocouple N 100 1300 C 148 2372 F Thermocouple L 100 900 C 148 1652 F Therm...

Page 9: ...0 see section 1 29 7 This pa rameter is associated to the specific model of probe installed Sensor Jumper S see section 1 9 Jumper T see section 1 9 Range in C in F Total error Connection terminals Cu...

Page 10: ...n for internal jump ers is indicated at section 1 9 See below connections for NTC probe 1 14 Measuring with NTC probes NTC R25 configurable Jumper S see section 1 9 Jumper T see section 1 9 Range of m...

Page 11: ...nnection terminals Zin 4 20 mA passive needs Vexc 0 100 0 de 9999 a 1999 D 1 2 0 15 25 mA 2 signal 5 Vexc 4 7Ohm active 4 5 2 mA 4 mA 0 10 Vdc passive needs Vexc A 1 2 0 20 25 Vdc 2 Vdc 4 comm 5 Vexc...

Page 12: ...n 1 9 for a list of available functions Start up connections and jumpers For instrument start up follow the steps listed at section 1 7 Signal connections are indicated at section 1 5 Location for int...

Page 13: ...ree potential contact called Exter nal control To configure the second scaling 1 configure the internal jumper T to activate the EK external control function see section 1 9 2 associate the external c...

Page 14: ...X and then insert clips A B C and D in this order The internal structure of the instrument is shown in the graphic be low Module Opt 2 connects to module Opt 1 Module Opt 1 connects to the display Opt...

Page 15: ...t validates the num ber displayed Key UP 5 the UP 5 key gives access to the fast access menu Inside the menus it moves vertically through the different menu op tions At menus with numerical value entr...

Page 16: ...able options Pt100 3 wire Pt100 2 wire Pt500 Pt1000 Ni100 Ni200 and Ni1000 For measures with 2 wires a manual compensation for the wire resistance error is available us ing parameter Offset reading oF...

Page 17: ...ximum type alarm or minimum type activates when the dis play value is higher or lower than the setpoint value at the Setpoint SEt parameter enter the value for the alarm activation point This paramete...

Page 18: ...vate a tare to the signal The actual signal will read 0 The calibration is not changed select parameter Decimal point 0 dP 0 to remove the decimal point on display select parameter Decimal point 1 dP...

Page 19: ...compen sate for the possible error introduced by the resistance of the signal wires when measuring Pt probes Ni probes or resistances For more information see section 1 22 1 29 7 Menu Tools Option 1...

Page 20: ...Ni1000 PTC and NTC For thermocouple probes resolution is fixed to 1 at the Degrees dEG menu select the temperature values to be read in C degrees Celsius or in F degrees Fahrenheit at the Alpha ALPh...

Page 21: ...r 485 at Tools out 1 r 485 to have the serial output configuration at this menu entry Configure the address for the instrument parameter Address 1 to 247 Addr the bus speed parameter Speed kbps bAud...

Page 22: ...c range 2 Vdc range 200 mVdc range 60 mVdc range 5 Adc range 20 mAdc range Process 4 20 mA range 0 10 Vdc range Thermocouples thermocouple K thermocouple J thermocouple E thermocouple N thermocouple L...

Page 23: ...ss Memory of max Memory of min Setpoint 1 Setpoint 2 Degrees degrees Celsius degrees Fahrenheit Temperature resolution 1 resolution 0 1 resolution Second scaling Decimal point Display Low Display High...

Page 24: ...elected at Tool Out 1 Analog output Display Low Display High Modbus RTU Address 1 to 247 Speed Kbps Format Password Version Factory reset Minimum Brightness Maximum Standard Alpha Alpha 385 Alpha 390...

Page 25: ...ent Industrial CISPR 11 Instrument Class A Class B Group 1 For a detailed declaration see document www fema es docs 4152_CE Declaration_C40 D_en pdf Barber del Vall s January 2017 Xavier Junc Product...

Page 26: ...portional to the reading of the instrument The mA output can be connected to work in active loop the module provides the power of the loop or passive mode the power of the loop is not provided by the...

Page 27: ...420 see section 1 29 7 to inform the instrument that there is a M1 analog output module installed From this point you will need the calibration sheet provided together with the module M1 This sheet s...

Page 28: ...the front filter Inter nal press buttons for configuration are accessible Optionally request the instrument preconfigured from factory Without front keypad 3 1 Option NBT O ring for IP65 front protec...

Page 29: ...ll mount housing Together with the KIP protector offers a full IP65 protection For Series C of panel meters 4 4 WME wall housing Benchtop housing for Series C and of panel meters Handle with three sel...

Page 30: ...FEMA ELECTR NICA Series C C40 D 30 Notes...

Page 31: ...FEMA ELECTR NICA Series C C40 D 31 Notes...

Page 32: ......

Reviews: