FEMA ELECTRÓNICA . Series S . S40-P
6
Decimal point
1.10 Configuration menu
Press ‘SQ’ (
<
) for 1 second to access the ‘
configuration menu’.
For a description on how to operate inside the menus see sec
-
tion
. For a full vision of the ‘
configuration menu
’ structure
1.11.
Scaling
Input low
Display low
Input high
Display high
To configure the initial set up of the instrument, select the input sig
-
nal range, the decimal point position, and scale the reading.
Access the ‘
Input
’ (‘
InP
’) menu to select the input signal range. Op
-
tions available are 4/20
mA, 0/10
Vdc, ±20
mA y ±10
Vdc. Select the
rear jumper at position ‘4’ for Vdc or ‘5’ for mA
Access the ‘
Decimal point
’ (‘
dP
’) menu to select the decimal point
position. Move the decimal point by pressing the ‘LE’ (
3
) key.
Scale the reading at the ‘
Scaling
’ (‘
ScAL
’) menu. There are four pa
-
rameters to configure :
• ‘
Input low
’ (‘
ILo
’)
- minimum input signal, in mA or Vdc with
two decimals.
• ‘
Display low
’ (‘
dLo
’) - reading associated to the minimum input
signal.
• ‘
Input high
’ (‘
IhI
’)
- maximum input signal, in mA or Vdc with
two decimals..
• ‘
Display high
’ (‘
dhI
’) - reading associated to the maximum input
signal.
The ‘
Field correction
’ (‘
F.cor
’) functions allows to correct the instru
-
ment scaling parameters, when the instrument is being installed on
the field, providing a fast and easy way to compensate offset signals.
To operate the ‘
Field correction
’ (‘
F.cor
’) function for the offset cor
-
rection, generate the minimum input signal and access the ‘
Field
low
’ (‘
F.Lo
’) function. Then the following sequence will start
• message with signal type (‘
mA
’ or ‘
Vdc
’)
• message ‘
wait
’ (‘
WAIt
’) in flash mode
• after 5 seconds, message ‘
ok
’ (‘
oK
’) appears
• at this point, press key SQ (‘
<
‘)
• the menu returns to menu entry ‘
Field Low
’ (‘
F.Lo
’)
The instrument has read the input signal and automatically config
-
ures the ‘
Input low
’ (‘
I.Lo
’) for you.
Repeat the process generating the maximum input signal and access
the ‘
Field high
’ (‘
F.hI
’) menu. The instrument will read the input sig
-
nal and automatically configure the ‘
Input high
’ (‘
I.hI
’) parameter to
match the signal read.
Application - a 0/10 Bar pressure transmitter provides a 4/20 mA out-
put signal. At installation, the operator detects that the reading at
0 Bar is 0.34 Bar and that at 10 Bar the reading is 10.72 Bar. Apply the
‘
Field correction
’ / ‘
F.Lo
’ function while reading is 0.34 Bar and the
instrument will automatically correct the reading to 0.00 Bar. After-
wards, apply the ‘
Field correction
’ / ‘
F.Hi
’ function while reading is
10.72 Bar and the instrument will automatically correct the reading
to 10.00 Bar.
1.10.1
Initial set-up
1.10.2
Field correction
Range de 4/20 mA
Range de 0/10 Vdc
Range de ±20 mA
Range de ±10 Vdc
Input
Field correction
Correction high
Correction low
Wait (5 sec.)
(
<
)
Wait (5 sec.)
(
<
)