User’s Manual S40-R

FEMA ELECTRÓNICA - Page 8

The “input” menu selects the input signal range. Options are  :

   Ratiometric signal

   Potentiometer <5K 

   Potentiometer <5M  

in this range the value of the poten-

tiometer must be introduced (in KOhm)

The display scaling is based on 4 parameters. These param-

eters define the two points of the straight line “signal / display”.

Input Low (“ILo”)   

Signal input low

Display Low (“dLo”) 

Display low

Input High (“IhI”) 

 

Signal input high

Display High (“dhI”) 

Display high

The instrument can manage up to 2 alarms. These alarms 
control optional relays R1 (see section 2.1) which can be in-
stalled at slots Opt1 and Opt2. More alarms can be achieved 

by installing special control modules R2 and R4. Configuration 

menus for special modules are not listed in this document.

Active (ACT) -

 Value “ON/OFF”. Defines if the instrument 

has to manage this alarm or not. Select “OFF” for alarm not 
managed. 

Type (TYPE) 

- Value “MAX/MIN”. Defines the behavior of the 

alarm as maximum or minimum alarm. The alarms configured 

as maximum are activated when the display value is equal or 

higher than the setpoint. The alarms configured as maximum 

are deactivated when the display is lower than the setpoint. 

The alarms configured as minimum have the inverse behavior.

Setpoint (SET) - 

Value from “9999” to “-1999”. Alarm set point.

Select the position for the decimal point. Modify with the LE 
button and validate with the SQ button.

Hysteresis (HYST) -

 Value from “0” to “9999”. Points of hyster-

esis. The hysteresis applies on the deactivation of the alarm.

Delay (DEL) -

 Value from “0.0” to “99.9” seconds. Delay to be 

applied to the relay activation and deactivation. Relays are 
activated and deactivated  X seconds after the activation / 
deactivation of the alarm. The delay affects only to the relays. 
The delay does not affect to the front leds.

Setpoint2 (SET2) - 

Value from  “-1999” to “9999”. Second 

setpoint. The second setpoint allows for the creation of acti-

vation windows. If the alarm is configured as maximum with 

setpoint 1000 and setpoint2 is configured at 1500, the alarm 

will be activated between 1000 and 1500 and the alarm will 
be deactivated when display is <1000 and  >1500. Setpoint2 
is affected on the same way as the setpoint with hysteresis 
and delays.

Functions on this menu allow for configuration of the display.

Fixed Digits (FIX.D) -

 Allows a digit to be fixed to a predefined 

value (for example, least significant digit fixed to “0”). It is a 

condition to fix a digit that all digits to its right are also fixed. 

Value “-” shows that the digit is not fixed.

Average (AVR) -

 Recursive filter applied to display. Value 

from “0.0” to “99.9”. The severity of the filter increases with 

the value selected. Increasing the severity of the filter makes 

the display response slower.

Steps (STEP) -

  Display changes on predefined steps. Values 

are 1, 2, 5, 10, 20 and 50. The display is made in steps of X 
counts. For example, select a step of 20 will make the display 
to change in steps of 20 (1420, 1440, 1460, ...).

Left Zero (LZER) - 

Value “On/Off”. Select “On” to visualize 

zeros to the left.

Maximum (MAX) -

 Memory of maximum display. Indicates 

the maximum value of display since the last reset of the 
memory. Memory is reset on the following cases : manual 

reset from the Configuration Menu (Maximum), change on 

the input signal (Input), modification on the scaling (Scaling), 

change on the decimal point (dP), modification of the linear

-

ization segments, or instrument power-down.

Minimum  (MIN) -

 Memory of minimum display. Indicates the 

minimum value of display since the last reset of the memory. 
Memory is reset on the following cases : manual reset from 

the  Configuration  Menu  (minimum),  change  on  the  input 

signal (Input), modification on the scaling (Scaling), change 

on the decimal point (dP), modification of the linearization 

segments, or instrument power-down.

Assigns the current input signal value to the Input Low or Input 

High configuration parameters. The new values for Input and 

Display are displayed.

Auto Low (A.LO) -

 Select “A.LO” to set the current input 

signal value to the Input Low configuration parameter.

Auto High (A.HI) -

 Select “A.HI” to set the current input signal 

value to the Input High configuration parameter.

1.10.1 Input menu

1.10.2 Scaling

1.10.3 Decimal point

1.10.4 Auto correction

1.10.5 Alarms

1.10.6 Display

Summary of Contents for S Series

Page 1: ...earization maximum and minimum display memory scalable reading filters steps signal autocorrection measure function peak hold alarms with double setpoints 5 levels of brightness Universal AC and DC po...

Page 2: ...dard IP54 front protection with optional upgrade to IP65 protection Connections via plug in screw terminals and configuration via three front push buttons For application on industrial environments S4...

Page 3: ...wer line a protection fuse acting as disconnection element easily accessible to the operator and identified as a protection device Power H fuse 250mA time lag Power L fuse 400mA time lag 6 7 8 9 0 1 2...

Page 4: ...ec 0 to 99 signal Power Power H 85 to 265 Vac dc Power L 11 to 60 Vdc and 24 48Vac Consumption 4W Isolation 3500Veff for power H 2000Veff for power L all levels tested for 60 seconds Configuration 3 f...

Page 5: ...ue outputs The Information Menu is for information only and it does not accept changes on the displayed information To enter the In formation Menu press the UP button It is not affected by the PASSWOR...

Page 6: ...ut High Display High Decimal Point Alarm1 Setpoint Hysteresis Delay Setpoint2 Counts Seconds Second setpoint Auto correction Auto High Auto Low Alarms Signal Low Display Low Signal High Display High R...

Page 7: ...zation Password Version Points Enter signal values Input and associated display val ues Display for each linearization point Then activate Act to On In case of error verify the points data Select pass...

Page 8: ...alarm will be activated between 1000 and 1500 and the alarm will be deactivated when display is 1000 and 1500 Setpoint2 is affected on the same way as the setpoint with hysteresis and delays Function...

Page 9: ...al while hold is active 1 10 7 Tools 1 10 8 Menu OptX Options When the instrument detects that the displayed value does not correspond to the expected value the display will flash and alternate with a...

Page 10: ...he configured input signal range 4 20mA 0 10Vdc and the values for input low ILo display low dLo input high IhI and display high dhI Maximum MAX Value of the maximum display Minimum MIn Value of the m...

Page 11: ...to the 61010 1 CE security regulation and is designed for applications on industrial environ ments See the CE Declaration of Conformity further in this document for information on the category of meas...

Page 12: ...AO modules can be installed in one S Series panel meter For more information see document 2657_S40_OPTIONAL_ MODULES at www fema es Module with 1 relay Up to a maximum of two R1 modules can be instal...

Page 13: ...contacts Common NO NC Maximum current 6A resistive load each relay Voltage 250 Vac continuously Installable at slot Opt 1 R2 fills OPT1 R4 fills OPT1 and OPT2 Terminal Plug in screw terminals pitch 3...

Page 14: ...User s Manual S40 R FEMA ELECTR NICA Page 14 T H I S P A G E B L A N K...

Page 15: ...User s Manual S40 R FEMA ELECTR NICA Page 15 T H I S P A G E B L A N K...

Page 16: ...other products ELECTRONIC INSTRUMENTATION FOR INDUSTRY www fema es FEMA ELECTR NICA S A Altimira 14 Pol Ind Santiga E08210 Barber del Vall s BARCELONA SPAIN Tel 34 93 729 6004 www fema es Fax 34 93 72...

Reviews: